Parameters:
<StatisticCount>
Number of measurement intervals for the "RF Carrier" measure-
ment
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Measurement Control Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
"RF Carrier / RF Modulation / AF > Statistic Count"
CONFigure:FMSTereo:MEAS<i>:MEValuation:SCOunt:RFModulation
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatisticCount>
Number of measurement intervals for the "RF Modulation" mea-
surement
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Measurement Control Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
"RF Carrier / RF Modulation / AF > Statistic Count"
CONFigure:FMSTereo:MEAS<i>:MEValuation:SCOunt:AF
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<StatisticCount>
Number of measurement intervals for the "AF Modulation" mea-
surement
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Measurement Control Settings