4.2.1 Signal Routing and Analyzer Settings
The measurement is configured using the parameters in the "FM Stereo Radio Config-
uration" dialog. The following parameters configure the RF input path.
See also: "Connection Control (Measurements)" in the R&S
CMW user manual, chap-
ter "System Overview"
Fig. 4-1: FM Stereo Radio Configuration: Signal routing and analyzer settings
Scenario
Selects the measurement scenario. The measurement can be used in "Standalone"
mode, using all "FM Stereo Radio" measurement settings.
Remote command:
ROUTe:FMSTereo:MEAS<i>:SCENario?
ROUTe:FMSTereo:MEAS<i>?
RF Routing
Selects the input path for the measured RF signal, i.e. the input connector and the RX
module to be used.
Depending on your hardware configuration there may be dependencies between both
parameters. Select the RF connector first. The "Converter" parameter offers only val-
ues compatible with the selected RF connector.
Remote command:
ROUTe:FMSTereo:MEAS<i>:SCENario:SALone
External Attenuation (Input)
Defines the value of an external attenuation (or gain, if the value is negative) in the
input path. The power readings of the R&S
CMW are corrected by the external attenu-
ation value.
The external attenuation value is also used in the calculation of the maximum input
power that the R&S
CMW can measure.
If a correction table for frequency-dependent attenuation is active for the chosen con-
nector, then the table name and a button are displayed. Press the button to display the
table entries.
Remote command:
CONFigure:FMSTereo:MEAS<i>:RFSettings:EATTenuation