The statistical values in the tables are calculated as follows:
●
Current:
Value of the result obtained in the last measurement interval.
●
Average:
Average of all "Current" values referenced to the last statistics cycle.
●
Max:
Largest "Current" value that the R&S
CMW obtained since the start of the
measurement.
●
Min:
Smallest "Current" value that the R&S
CMW obtained since the start of the
measurement.
●
StdDev:
Standard deviation of all "Current" values since the start of the measure-
ment.
All statistical results (statistical tables and "Average" or "Max" traces) are calculated
according to the general rules for statistical results.
See also: "Statistical Results" in the R&S
CMW user manual, chapter "System Over-
view"
Statistic Count
Progress bar for the measurement, displayed in all detailed views. During the first sin-
gle shot after the start of the measurement, the bar shows the number of elapsed mea-
surement intervals relative to the "Statistic Count". A filled progress bar indicates that
the first shot is complete and the statistical depth has been reached.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "System Over-
view"
Out of Tolerance
Percentage of measurements that were failed because they exceeded the specified
limits.
Measurement Results