Operating the Instrument
R&S
®
FSW
91
Getting Started 1338.4102.02 ─ 08
7.1.3
Marker Information
Marker information is provided either in the diagram grid or in a separate marker
table, depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last two markers or
delta markers that were set are displayed, if available, as well as their index. The
value in the square brackets after the index indicates the trace to which the
marker is assigned. (Example: M2[1] defines marker 2 on trace 1.) For more than
two markers, a separate marker table is displayed beneath the diagram by
default.
Marker information in marker table
In addition to the marker information displayed within the diagram grid, a separate
marker table may be displayed beneath the diagram. This table provides the fol-
lowing information for all active markers:
Type
Marker type: N (normal), D (delta), T (temporary, internal), PWR
(power sensor)
Ref
Reference (for delta markers)
Trc
Trace to which the marker is assigned
X-value
X-value of the marker
Y-value
Y-value of the marker
Func
Activated marker or measurement function
Func .Result
Result of the active marker or measurement function
The functions are indicated with the following abbreviations:
FXD
Fixed reference marker
PHNoise
Phase noise measurement
CNT
Signal count
TRK
Signal tracking
NOIse
Noise measurement
Understanding the Display Information