Command Reference
R&S
®
ZNB/ZNBT
974
User Manual 1173.9163.02 ─ 53
CONTrol<Ch>:RFFE<Bus>:TEST:DATA:RANGe
<Data Current Range>
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk:RANGe
<Clock Current Range>
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:RANGe
<CurrentRange>
Defines the (upper bound of the) current range for the voltage/current measurement on
the respective RFFE pin. The analyzer firmware automatically selects a suitable shunt
resistance, which can be queried using
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:
.
Suffix:
<Ch>
.
Channel number
<Bus>
RFFE bus interface 1 or 2
Parameters:
<CurrentRange>
Range:
2 x 10
n
μA with n=1,...,5
Default unit: A
Example:
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:SHUNt?
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk:SHUNt?
CONTrol<Ch>:RFFE<Bus>:TEST:DATA:SHUNt?
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:SHUNt?
Returns the shunt resistance (in Ω) selected by the analyzer firmware for the config-
ured current range (see
CONTrol<Ch>:RFFE<Bus>:TEST:VIO:RANGe
).
The dependency between current range and shunt resistance is displayed in the table
below.
Suffix:
<Ch>
.
Channel number
<Bus>
RFFE bus interface 1 or 2
Example:
:CONTrol1:RFFE1:TEST:DATA:RANGe 20.0mA
:CONTrol1:RFFE1:TEST:CLOCk:RANGe 2mA
:CONTrol1:RFFE1:TEST:VIO:RANGe 0.2mA
:CONTrol1:RFFE1:TEST:DATA:SHUNt?
Returns 10 [Ω]
:CONTrol1:RFFE1:TEST:VIO:RANGe 2mA
:CONTrol:RFFE1:TEST:CLOCk:SHUNt?
Returns 100 [Ω]
:CONTrol1:RFFE1:TEST:VIO:RANGe 0.2mA
:CONTrol:RFFE1:TEST:VIO:SHUNt?
Returns 1000 [Ω]
Usage:
Query only
Options:
R&S ZN-B15/Z15 Var. 03
SCPI Command Reference