Getting Started
R&S
®
ZNB/ZNBT
79
User Manual 1173.9163.02 ─ 53
VNA
1
DUT
1
You can also use the basic transmission test setup, e.g. if you want to measure
reflection and transmission parameters in parallel.
●
The analyzer provides special calibration types for reflection measurements. Use
the calibration wizard and select an appropriate type.
A full n-port calibration (TOSM, UOSM, TNA ...) corrects the system errors for all
transmission and reflection S-parameters.
●
Some of the trace formats are particularly suited for reflection measurements. For
instance, you can display the measured reflection coefficient S
11
in a Smith chart to
obtain the complex input impedance at port 1.
Chapter 4.1.7, "Starting the Analyzer and Shutting Down"
on page 23 to shut down your analyzer.
Performing Measurements