GUI Reference
R&S
®
ZNB/ZNBT
553
User Manual 1173.9163.02 ─ 53
Access:
CHANNEL – [CHANNEL
CONFIG] > "Mixer Mode" > "Mixer Meas Wizard..."
The setup of the mixer measurement proceeds in 4 steps:
1. Define the mixer.
Chapter 6.12.6.3, "(Mixer) Setup Dialog"
2. Select the mixer parameters to be measured. Depending on your selection, the
required channels and traces are set up automatically; the active channel and its
first trace will be renamed, all other channels and traces in the active recall set will
be deleted.
Table 6-3: Auto-created channels and traces
Measurement Parameter
Channel Name
Trace Name
Conversion Gain S<j><i>
Ch_M
Conv
RF Reflection S<i><i>
Ch_M
RF_Refl
IF Reflection S<j><j>
Ch_M
IF_Refl
RF Isolation S<j><i>
1)
Ch_RF
RF_Isol
LO[m] Leakage S<i><k>
Ch_LO
LO_Leak
Channel Config Softtool