
Command Reference
R&S
®
ZNB/ZNBT
1252
User Manual 1173.9163.02 ─ 53
Parameters:
<ReduceAt
SweepEnd>
OFF
(default): at sweep end the output power of the sweep start
is restored
ON
: at sweep end the output power of the driving port is reduced
as if the channel base power was set to its minimum possible
value
Example:
See
REDuce:STATe OFF
is equivalent to
SWEepend:MODE AUTO
REDuce:STATe ON
is equivalent to
SWEepend:MODE REDuce
SOURce<Ch>:POWer<PhyPt>:STATe
<Boolean>
Turns the RF source power at a specified test port on or off.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
Parameters:
<Boolean>
ON | OFF - turns the internal source power at the specified test
port no.
<PhyPt>
on or off.
*RST:
ON
Example:
*RST; SOUR:POW -6
Set the internal source power for channel 1 and all test ports to
-6 dBm.
SOUR:POW2:STAT?
Query whether the source power at test port 2 is on. The ana-
lyzer returns a
1
.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:STARt
<StartPower>
SOURce<Ch>:POWer<PhyPt>:STOP
<StopPower>
These commands define the start and stop powers for a power sweep. The values also
define the left and right edges of a Cartesian diagram. A power sweep must be active
(
POWer
) to use these commands.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected power sweep range applies to all source ports used
in the active channel.
SCPI Command Reference