SLIT LAMP RSL 4300/4500 by RODENSTOCK

SLIT LAMP RSL 4300/4500 by RODENSTOCK

The manufacturer declines any and all responsibility and warranty 

coverage should the instrument be tampered with in any manner or 

should routine maintenance be omitted or performed in manners not in 

accordance with these manufacturer’s instructions.
There are no user serviceable parts in this instrument. Any servicing 

or repairs should only be carried out by Rodenstock Instruments or by 

suitably trained and authorised distributors. Service manuals will be 

available to authorised Rodenstock service centres and Rodenstock 

trained service personnel.

9.  SPECIFICATIONS AND ELECTRICAL RATINGS

The Rodenstock Slit Lamp is a medical electrical instrument. The instrument requires special 

care concerning electromagnetic compatibility (EMC). This Section describes its suitability 

in terms of electromagnetic compatibility of this instrument. When installing or using this 

instrument, please read carefully and observe what is described here.
Portable or mobile-type radio frequency communication units may have an adverse effect on 

this instrument, resulting in malfunctioning.

9.1  ELECTROMAGNETIC EMISSIONS

Guidance and manufacturer’s declaration – electromagnetic emissions 

The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specifi ed 

below. The customer or user should ensure that it is used in such an environment.

Emissions test

Compliance Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The Rodenstock Slit Lamp uses RF energy only for its 

internal function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

RF emissions CISPR 11

Class B

The Rodenstock Slit lamp is suitable for use in a 

professional healthcare facility environment. The 

Rodenstock Slit lamp is not intended for use in home 

environment.

Harmonic emissions 

IEC 61000-3-2

Class B

Voltage fl uctuations / 

fl icker emissions IEC 

61000-3-3

Complies

9.2  ELECTROMAGNETIC IMMUNITY

Guidance and manufacturer’s declaration – electromagnetic immunity

The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specifi ed 

below. The customer or user should ensure that it is used in such an environment.

Immunity test IEC 60601 Test

level

Compliance 

level

Electromagnetic 

environment – guidance

Electrostatic 

discharge (ESD).
IEC 6100-4-2

± 8 kV contact
± 15 kV air

± 8 kV contact
± 15 kV air

Floors should be wood, concrete 

or ceramic tile. If fl oors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%

Electrical fast 

transient/burst. 
IEC 61000-4-4

± 2 kV for power 

supply lines
± 1 kV for power  

supply lines

± 2 kV for power 

supply lines
N/A

Mains power quality should be 

that of a typical professional 

healthcare facility

Surge.
IEC 61000-4-5

± 1 kV line(s) to 

line(s)
± 2 kV line(s) for 

input/output line(s)

± 1 kV line(s) to 

line(s)
± 2 kV line(s) for 

input/output line(s)

Mains power quality should be 

that of a typical professional 

healthcare facility

Voltage dips, short 

interruptions and 

voltage variations 

on power supply 

input lines.
IEC 61000-4-11

U

T

 = 0% 0.5 cycle

(0, 45, 90, 135, 180,
225, 270, 315°)

U

T

 = 0%; 1 cycle

U

T

 = 70%;

25/30 cycles (@ 0°)

U

T

 = 0%;

250/300 cycle

U

T

 = 0% 0.5 cycle

(0, 45, 90, 135, 180,
225, 270, 315°)

U

T

 = 0%; 1 cycle

U

T

 = 70%;

25/30 cycles (@ 0°)

U

T

 = 0%;

250/300 cycle

Mains power quality should be 

that of a typical professional 

healthcare facility environment. 

If the user of the Rodenstock 

Slit Lamp requires continued 

operations during power mains 

interruptions, it is recommended 

that the instrument be powered 

from an uninterruptible power 

supply.

Power frequency 

(50/60 Hz)
Magnetic fi eld. IEC 

61000-4-8

30 A/m

30 A/m

Power frequency magnetic fi elds 

should be at a level characteristic 

of a typical location in a typical 

professional healthcare facility 

environment.

Note: 

U

T

 is the a. c. mains voltage prior to application of the test level.

Immunity 

test

IEC 60601
Test level

Compliance 

level

Electromagnetic environment 

– guidance

Portable and mobile RF communications equipment 

should be used no closer to any part of the 

Rodenstock Slit Lamp, including cables, than the 

recommended separation distances calculated from 

the equation applicable to the frequency of the 

transmitter.

Содержание RSL 4300

Страница 1: ...RSL 4300 4500 SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...TRUCTIONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYST...

Страница 3: ...s the eyelid sclera conjunctiva iris natural crystalline lens and cornea The binocular slit lamp provides stereoscopic magnified view of the eye structures in detail enabling anatomical diagnoses to b...

Страница 4: ...ical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair After removal of the LED do not touch the Slit Lamp LED electrical contac...

Страница 5: ...FECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do...

Страница 6: ...to an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the co...

Страница 7: ...tor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 5 4 FITTING APPLANATION TONOMETERS T TYPE AND R TYPES Rodenstock Applanation Tonometer T...

Страница 8: ...ces by minus one 1 or minus two 2 dioptres 16 27 28 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with...

Страница 9: ...f an LED failure please contact Rodenstock or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean as per section 3 on...

Страница 10: ...ECTROMAGNETIC IMMUNITY Guidance and manufacturer s declaration electromagnetic immunity The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specified below The customer or...

Страница 11: ...le and mobile RF communications equipment and the Rodenstock Slit Lamp The Rodenstock Slit Lamp is intended for the use in an electromagnetic environment in which radiated RF disturbances are controll...

Страница 12: ...nnected to the power supply together the constitute a Medical Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti s...

Страница 13: ...V 3020 P 5040 ROD USB3 Type A to Micro B Cable Assy 3020 P 7107 XY Pad 405 EP39 80169 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the produ...

Страница 14: ...00 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51519 Issue 5 Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Stra e 21 90427 N rnberg Germany...

Отзывы: