SLIT LAMP RSL 4300/4500 by RODENSTOCK

SLIT LAMP RSL 4300/4500 by RODENSTOCK

3.  Attach the breath shield as indicated in the images below.

5.3  CABLE ATTACHMENT PROCEDURE

1.  Connect the power cables.

a)  Chinrest fi xation light cable to power supply unit.
b)  (3 pin) cable from power supply unit to slit lamp base assembly.
c)  (4 pin) main lamp cable from bottom of the illumination assembly to the slit lamp 

base assembly.

d)  Ensure cables are routed to allow free movement of the XYZ base and to be clear of 

the patients.

If your Slit Lamp was not supplied with a transformer (Part #3020-P-5040), 

make sure that the power connection is compatible with the specifi cations 

in this manual and is connected by a qualifi ed technician to an available 

and suitable power supply, see section 9.4 Power Supply page 20.

2.  If the Rodenstock Slit Lamp is used with a power supply or cables other than those 

supplied, this may result in increased emissions or decreased immunity of the 

Rodenstock Slit Lamp in relation to EMC performance. 

3.  Connect the mains power to the Slit Lamp transformer using the power lead provided.

4 pin
3 pin

3 pin to 3 pin

Fixation light socket

4.  To isolate from the mains supply unplug the mains plug from the wall socket. Ensure 

that the product is positioned so that it is easily accessible.

Only a hospital grade 3-conductor electrical power supply cable must be 

used. For USA and Canada: Detachable power supply cord set, UL listed, 

type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWG. Plug, cable 

and ground lead connection of the socket have to be in perfect condition.

5.4  FITTING APPLANATION TONOMETERS, T TYPE AND R TYPES

Rodenstock Applanation Tonometer (T-TYPE)

1.  Position the guide plate in the tonometer/test bar support hole on the slit lamp.
2.  Lift Tonometer out of the packaging and assemble it by inserting the pin on its base 

into one of the two possible openings (for right or left eye) on the horizontal guide 

plate above the slit lamp axis. These positions are related to the microscope optics and 

observation can be made either through the right or the left eye-piece.

The tonometer will slip easily onto the support plate; stability is assured by the locking 

pins.

3.  To obtain an image as clear and as free of refl exes as possible, the angle between the 

illumination and the microscope should be about 60° and the slit diaphragm should be 

fully opened.

4.  When not in use the Tonometer should be removed from the Slit Lamp and placed 

securely back in the packaging or a suitable location.

Applanation Tonometer ‘Rodenstock Fixed’ (R-TYPE)

This instrument is for those who wish the tonometer to remain permanently on the slit lamp.

5.  Mount the plate for the tonometer onto the microscope body using the securing screw.
6.  Mount the Tonometer onto the mounting post ensuring the grub screws are loose.
7.  Swing the Tonometer arm forward in front of the microscope for examination. Rotate 

the Tonometer body until the prism is in the centre of the view through the eyepiece. 

Carefully tighten the two retaining grub screws in turn until the Tonometer is secure on 

the mount and the prism is in the centre of view.

Right eye
Left eye

Содержание RSL 4300

Страница 1: ...RSL 4300 4500 SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...TRUCTIONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYST...

Страница 3: ...s the eyelid sclera conjunctiva iris natural crystalline lens and cornea The binocular slit lamp provides stereoscopic magnified view of the eye structures in detail enabling anatomical diagnoses to b...

Страница 4: ...ical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair After removal of the LED do not touch the Slit Lamp LED electrical contac...

Страница 5: ...FECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do...

Страница 6: ...to an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the co...

Страница 7: ...tor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 5 4 FITTING APPLANATION TONOMETERS T TYPE AND R TYPES Rodenstock Applanation Tonometer T...

Страница 8: ...ces by minus one 1 or minus two 2 dioptres 16 27 28 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with...

Страница 9: ...f an LED failure please contact Rodenstock or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean as per section 3 on...

Страница 10: ...ECTROMAGNETIC IMMUNITY Guidance and manufacturer s declaration electromagnetic immunity The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specified below The customer or...

Страница 11: ...le and mobile RF communications equipment and the Rodenstock Slit Lamp The Rodenstock Slit Lamp is intended for the use in an electromagnetic environment in which radiated RF disturbances are controll...

Страница 12: ...nnected to the power supply together the constitute a Medical Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti s...

Страница 13: ...V 3020 P 5040 ROD USB3 Type A to Micro B Cable Assy 3020 P 7107 XY Pad 405 EP39 80169 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the produ...

Страница 14: ...00 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51519 Issue 5 Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Stra e 21 90427 N rnberg Germany...

Отзывы: