SLIT LAMP RSL 4300/4500 by RODENSTOCK

SLIT LAMP RSL 4300/4500 by RODENSTOCK

Consult instructions for use

General warning sign

Date of manufacture

Warning: Electricity

Manufacturer’s name and 

address

Warning: Floor level obstacle

Country of manufacture

Warning: Non-ionizing 

radiation

Waste Electrical and Electronic 

Equipment (WEEE) recycling

Warning: Optical radiation

This way up

Warning: Hot surface

Keep dry

Conformité Européene

Fragile

Type B applied part

Do not use if package is 

damaged

Class II equipment

Temperature limit

Atmospheric pressure 

limitation

Authorised representative in 

the European Community

Humidity limitation

Catalogue number

Serial number

Translation

Medical device

The Rodenstock Slit Lamp is designed and built in conformity with Directive 93/42/EEC, Regulation (EU) 2017/745 

and ISO 13485 Medical Devices Quality Management Systems.
Classifi cation:  CE:  Class I
 

FDA: Class II

The information contained within this manual must not be reproduced in whole or part without the manufacturer’s 

prior written approval. As part of our policy for continued product development we the manufacturer reserve the right 

to make changes to specifi cations and other information contained in this document without prior notice.
This IFU is also available on the Rodenstock website.
Copyright © Rodenstock Instruments 2021. Published in the UK 2021.

1.  INDICATIONS FOR USE 

This device is intended to be used only by suitably trained and authorised healthcare 

professionals.

CAUTION: Federal Law restricts this device to sale by or on the order of a 

physician or practitioner.

Intended use / purpose of instrument

The Rodenstock Slit Lamp facilitates an examination of the anterior segment, or frontal 

structures and posterior segment, of the human eye, which includes the eyelid, sclera, 

conjunctiva, iris, natural crystalline lens and cornea. The binocular slit-lamp provides 

stereoscopic magnifi ed view of the eye structures in detail, enabling anatomical diagnoses to 

be made for a variety of eye conditions. 

Brief description of the instrument

This Rodenstock Slit Lamp can either be mounted onto a custom table top supplied by Rodenstock 

or can be mounted on a third party’s table top (refraction unit) by suitably trained technicians.
The Rodenstock Slit Lamp consists of 5 assemblies; Illumination Tower; Observation System; XYZ 

Translation Base; Chinrest Assembly and a Table Top with Power Supply and Accessory Drawer.
The light intensity is controlled by a variable rheostat located on the XYZ Translation Base. 

There are a number of selectable fi lters allowing the user to control the characteristics of the 

examination light. Many ophthalmic instruments illuminate the eye with intense light. The light 

intensity on the Rodenstock Slit Lamp is continuously adjustable from maximum to zero. 

2.  SAFETY

2.1  PHOTOTOXICITY

CAUTION: The light emitted from this instrument is potentially hazardous. 

The longer the duration of exposure, the greater the risk of ocular 

damage. Exposure to light from this instrument when operated at 

maximum intensity will exceed the safety guideline after 81 seconds.
While no acute optical radiation hazards have been identifi ed for slit 

lamps, we recommend keeping the intensity of the light reaching the 

patient’s retina to the minimum possible for the respective diagnosis. 

Children, people with aphakia and people suffering from eye conditions 

are most at risk. An increased risk may also occur if the retina is exposed 

to the same or a similar device with a visible light source within 24 hours. 

This applies, in particular, if the retina has been photographed with a 

fl ashbulb in advance. 
Rodenstock Instruments shall on request, provide the user with a graph 

showing the relative spectral output of the instrument.

Содержание RSL 4300

Страница 1: ...RSL 4300 4500 SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...TRUCTIONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYST...

Страница 3: ...s the eyelid sclera conjunctiva iris natural crystalline lens and cornea The binocular slit lamp provides stereoscopic magnified view of the eye structures in detail enabling anatomical diagnoses to b...

Страница 4: ...ical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair After removal of the LED do not touch the Slit Lamp LED electrical contac...

Страница 5: ...FECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do...

Страница 6: ...to an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the co...

Страница 7: ...tor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 5 4 FITTING APPLANATION TONOMETERS T TYPE AND R TYPES Rodenstock Applanation Tonometer T...

Страница 8: ...ces by minus one 1 or minus two 2 dioptres 16 27 28 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with...

Страница 9: ...f an LED failure please contact Rodenstock or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean as per section 3 on...

Страница 10: ...ECTROMAGNETIC IMMUNITY Guidance and manufacturer s declaration electromagnetic immunity The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specified below The customer or...

Страница 11: ...le and mobile RF communications equipment and the Rodenstock Slit Lamp The Rodenstock Slit Lamp is intended for the use in an electromagnetic environment in which radiated RF disturbances are controll...

Страница 12: ...nnected to the power supply together the constitute a Medical Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti s...

Страница 13: ...V 3020 P 5040 ROD USB3 Type A to Micro B Cable Assy 3020 P 7107 XY Pad 405 EP39 80169 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the produ...

Страница 14: ...00 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51519 Issue 5 Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Stra e 21 90427 N rnberg Germany...

Отзывы: