SLIT LAMP RSL 4300/4500 by RODENSTOCK

SLIT LAMP RSL 4300/4500 by RODENSTOCK

6.3  DESCRIPTION OF FILTERS, APERTURES AND MAGNIFICATIONS

Stereo microscope

Eyepieces 12.5x 
Dioptric adjustment 

+/– 8D

PD range 8° converging eyepieces   49mm – 77mm 
PD range 0° parallel eyepieces 

38mm – 85mm 

Convergent angle of optical axis 

13°

Magnifi cation change

5 step drum

3 step drum

Magnifi cation

Field of view

3

5

6x

34mm

3

3

10x

22mm

3

3

16x

14mm

3

3

25x

8.5mm

3

5

40x

5.5mm

Filters

a)  Clear
b)  Neutral Density
c)  Red Free
d)  Blue

Apertures

Aperture diameters (mm)

a

b

c

d

0.2

1

2

3

5

9

14 Reserved for

future use

7.  ROUTINE MAINTENANCE

The maintenance outlined below should only be carried out once the 

main power cable has been disconnected. If you have any problems that 

are not covered by the procedures described below, contact Rodenstock 

Instruments or your local supplier.

7.1  LED SYSTEMS

1.  LED’s typically have a life exceeding 10,000 hours of continuous use and therefore can 

be considered as a non-consumable item that will not require changing by a user.

2.  Whilst this is a signifi cant life expectancy, we suggest that the Slit Lamp is always 

switched off between examinations to conserve energy and LED life.

3.  In the unlikely event of an LED failure please contact Rodenstock or your local distributor 

for guidance on the replacement procedure.

7.2  REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT

1.  Routinely clean as per section 3 on page 6 cleaning instructions.

7.3  CLEANING THE ILLUMINATION PRISM

1.  The prism should be cleaned with a soft, clean lens cloth.
2.  Care must be taken to keep the objective and the eyepiece lenses clean – use only soft, 

clean lens cloths to clean optical surfaces.

7.4  ELECTRICAL CONNECTIONS

1.  Routinely check all electrical connections, cables and connectors. 

7.5  OPTICS

1.  The optics should be wiped clean of any loose dirt or debris with a suitable dust brush 

then cleaned with a soft dry lens cloth, washed linen or other non abrasive lens cleaning 

material.

7.6  AXLE AND MECHANICAL PARTS

If the Slit Lamp becomes hard to move on the glide plate, the plate should be cleaned with a 

lightly oiled cloth or silicon polish. The axle should be cleaned only with dry lint free cloths.

8.  WARRANTY

The Rodenstock Slit Lamps are guaranteed for two years against faulty workmanship materials 

or factory assembly. Warranty is on a Return To Base (RTB) basis at the cost of the customer 

and may be void if the Slit Lamp has not been regularly serviced.
The manufacturer’s warranty and terms and conditions are detailed on the Rodenstock website.
The mirror, main illumination lamp and general ‘wear and tear’ are excluded from our standard 

warranty.

Содержание RSL 4300

Страница 1: ...RSL 4300 4500 SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...TRUCTIONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYST...

Страница 3: ...s the eyelid sclera conjunctiva iris natural crystalline lens and cornea The binocular slit lamp provides stereoscopic magnified view of the eye structures in detail enabling anatomical diagnoses to b...

Страница 4: ...ical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair After removal of the LED do not touch the Slit Lamp LED electrical contac...

Страница 5: ...FECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do...

Страница 6: ...to an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the co...

Страница 7: ...tor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 5 4 FITTING APPLANATION TONOMETERS T TYPE AND R TYPES Rodenstock Applanation Tonometer T...

Страница 8: ...ces by minus one 1 or minus two 2 dioptres 16 27 28 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with...

Страница 9: ...f an LED failure please contact Rodenstock or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean as per section 3 on...

Страница 10: ...ECTROMAGNETIC IMMUNITY Guidance and manufacturer s declaration electromagnetic immunity The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specified below The customer or...

Страница 11: ...le and mobile RF communications equipment and the Rodenstock Slit Lamp The Rodenstock Slit Lamp is intended for the use in an electromagnetic environment in which radiated RF disturbances are controll...

Страница 12: ...nnected to the power supply together the constitute a Medical Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti s...

Страница 13: ...V 3020 P 5040 ROD USB3 Type A to Micro B Cable Assy 3020 P 7107 XY Pad 405 EP39 80169 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the produ...

Страница 14: ...00 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51519 Issue 5 Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Stra e 21 90427 N rnberg Germany...

Отзывы: