background image

 20

 21

EN

EN

SLIT LAMP RSL 4300/4500 by RODENSTOCK

SLIT LAMP RSL 4300/4500 by RODENSTOCK

Horizontal fine adjustment

12mm

Table top dimensions

405 x 500mm

Fixation lamp

LED

Light source

LED

LED output power

240KLux (+/-20%)

Weight, packed  (approx.)

Slit Lamp complete

25Kg, 90 x 58 x 45cm W x D x H

Protection against ingress 

IPxO

Class II ME equipment

Insulation between mains parts and the functional earth provide at least two means of protection.

Power Supply 

Power supply unit

Switch mode, (100V-240V input) +/- 10% multi plug compliant to EN/

IEC 60601-1 EN/IEC 61000-6-2, EN/IEC 61000-6-3

Fuse

T2.5AH, 250V

Power supply output

12V DC: 2.5 amps must be EN/IEC 60601 compliant

Complies with

Electrical Safety (Medical) EN/IEC 60601-1
Electromagnetic compatibility EN/IEC 60601-1-2
Ophthalmic instruments - Fundamental requirements and test methods 

ISO 15004-1
Ophthalmic instruments - Optical radiation hazard ISO 15004-2

When the Slit Lamp is connected to the power supply together the constitute a Medical 

Electrical System as defined in EN/IEC 60601-1:2006.
The power supply forms a part of ME equipment.

Fuse Ratings and Quantity

2.5 amp anti-surge  

Fuse current 2.5A  

Voltage rating V AC 250V 

Breaking capacity 1500Amps  

Blow characteristic: Time Delay

Environmental Conditions:

USE

35

°C

10

°C

90%

30%

1060 hPa

800 hPa

Shock (without packing)

10 g, duration 6 ms

STORAGE CONDITIONS

*50

°C

-10

°C

95%

10%

1060 hPa

700 hPa

TRANSPORT CONDITIONS

*50

°C

-40

°C

95%

10%

1060 hPa

500 hPa

Vibration, sinusoidal

10 Hz to 500 Hz: 0.5g

Shock

30 g, duration 6 ms

Bump

10 g, duration 6 ms

*This instrument does not meet the temperature requirements of ISO 15004-1 for storage and transportation. Do 

not store or transport this instrument in conditions where the temperature may rise above 50 °C.

10. ACCESSORIES AND SPARES

Item

Part Number

1.27mm A/F ALLEN KEY

EP79-00766

1m DSL to PSU Cable

3020-P-7068

5mm L-Shaped Hexagonal Key

EP79-10892

6mm L-Shaped Hexagonal Key

EP79-10905

Black Patient Handle

EP79-80070

Cable Clip

EP39-80329

Chin Rest Pins (406)

EP39-80321

Digital Base to Computer Cable

3020-P-7077

Содержание RSL 4300

Страница 1: ...RSL 4300 4500 SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...TRUCTIONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYST...

Страница 3: ...s the eyelid sclera conjunctiva iris natural crystalline lens and cornea The binocular slit lamp provides stereoscopic magnified view of the eye structures in detail enabling anatomical diagnoses to b...

Страница 4: ...ical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair After removal of the LED do not touch the Slit Lamp LED electrical contac...

Страница 5: ...FECTION INSTRUCTIONS Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Only manual non immersion cleaning as described should be used for this instrument Do...

Страница 6: ...to an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the co...

Страница 7: ...tor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 5 4 FITTING APPLANATION TONOMETERS T TYPE AND R TYPES Rodenstock Applanation Tonometer T...

Страница 8: ...ces by minus one 1 or minus two 2 dioptres 16 27 28 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with...

Страница 9: ...f an LED failure please contact Rodenstock or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean as per section 3 on...

Страница 10: ...ECTROMAGNETIC IMMUNITY Guidance and manufacturer s declaration electromagnetic immunity The Rodenstock Slit Lamp is intended for use in the electromagnetic environment specified below The customer or...

Страница 11: ...le and mobile RF communications equipment and the Rodenstock Slit Lamp The Rodenstock Slit Lamp is intended for the use in an electromagnetic environment in which radiated RF disturbances are controll...

Страница 12: ...nnected to the power supply together the constitute a Medical Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti s...

Страница 13: ...V 3020 P 5040 ROD USB3 Type A to Micro B Cable Assy 3020 P 7107 XY Pad 405 EP39 80169 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the produ...

Страница 14: ...00 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51519 Issue 5 Date of Issue 12 05 2021 SUPPLIER Rodenstock Instruments Business Unit of Tomey GmbH Wiesbadener Stra e 21 90427 N rnberg Germany...

Отзывы: