Administrator Functions
Adjustment of the Image Analysis using the Live Operator
7.5.7.
Working with the Live Operator
The
Live Operator
window can be opened via the
Cell Type List
window, the
Cell Type
window and the
Measurement
window. If the Live Operator button in a
Measurement
window containing specifi c results
is clicked on, the
Live Operator
will open with the images of that measurement already loaded. If the
Live
Operator
is opened via the
Cell Type
or
Cell Type List
windows, the fi rst measurement in the
Measurement
List
will be loaded.
In the
Live Operator
window, some of the 11 parameters can be displayed on the screen at the same time.
Parameters at fi rst not visible can be scrolled into view using the scroll bar under the Operator Parameter
display. The table below displays the positions and settings of the 11 Operator parameters. See also “
Description
of the Individual Operator Parameters
” for a detailed description of each parameter.
Identifi er
Minimum
Maximum
Min. Size
Std. Size
Max. Size
Classifi cationBias
-20
20
-1
0
-8
CMinSize [µm]
0
100
3
7
10
LCAppearance
-10
10
4
0
0
DCAppearance
-10
10
0
0
-1
CMaxSize [µm]
0
100
70
70
100
AggrMinSize [µm]
8
50
12
12
15
BorderWidth
0
1000
10
10
10
AggrMaxSize [µm]
10
200
100
100
100
LCAggrApp
-10
10
5
0
0
DCAggrApp
-10
10
6
0
-3
MinPollutionArea [µm]
50
1000
100
100
100
Table 1: A list of the Operator parameters