Cedex HiRes Analyzer Operator’s Guide
D
D
Administrator Functions
Adjustment of the Image Analysis using the Live Operator
After loading a measurement (Reactor ID and Sample ID) and selecting an image, the selected image is displayed
in the upper right-hand corner of the
Live Operator
window. Each single cell detected by Cedex HiRes Analyzer
is marked as dead (red circle) or living (green circle).
Figure 75: Setting parameters for Cell Type Standard Cells
In the
Process Data
area, the values for new and previous results are displayed for viable cells (the number of
living cells of the current image) dead cells (the number of dead cells of the current image) and %viability.
If an Operator parameter is modifi ed, the newest values will appear under
New
column when an automatic
reprocessing of the selected image is fi nished. The values of the previous processing are displayed under
Previous
column. Thus, direct comparison to each previous evaluation is possible.
The labeling of the cells can be switched on and off in the upper-right hand corner in order to allow for better
evaluation of the results from changes in parameter settings. If multiple parameter changes should be made at
the same time, remove the check-mark in the
Automatic process
box by clicking on it. When the check-mark
is removed, the button
Process now
becomes active. Make all changes then click on the
Process now
button for
an analysis based on the new parameter settings.