The code on the test strip does not match the code in the
meter.
Recode the meter with the code strip from the pack
of test strips you are using (see Section 3.1).
The test strip is not suitable for the meter.
(e.g. a BM-Lactate test strip was used)
Use only Accutrend
®
Glucose or Accutrend
®
Cholesterol test strips.
The reading in of the code has failed.
Repeat the coding procedure (see Section 3.1).
The test strip was not properly inserted.
Open the flap, take out the strip, close the flap and
insert the strip again.
The flap was open during measurement.
Repeat the entire measuring procedure with a new
test strip and with the flap closed (see Section 3.2).