Batteries are running low.
(Does not apply during the function test)
About 50 measurements are still possible after the
first appearance of
LOW BAT
.
Have new batteries ready. Always change all three
batteries at the same time (see Section 5.3).
It is either too cold or too warm for a measurement.
All chemical reactions are temperature-dependent.
The same applies to all test strips. The admissible
temperature range is +18°C to +32°C for
Accutrend
®
Glucose and +18°C to +30°C for
Accutrend
®
Cholesterol. Accutrend
®
GC offers an
additional safety warning: if a measurement is per-
formed outside the above temperature ranges the
warning
TTT
flashes alternately with the result. The
reading may only be used for orientation.
Let the meter and test strips adjust slowly to room
temperature. Repeat the measurement within the
admissible temperature range (see Section 5.4).