Beacon 110 Rig Monitor Operator’s Manual
34
Buzzer Not
Working
•
The buzzer does not
sound an audible
alarm during alarm
conditions.
•
The buzzer sounds
weak or broken.
•
The buzzer is
disconnected or
misconnected.
•
The buzzer is
malfunctioning.
1. Verify that the wiring to the
ALARM
&
-
terminals of the
Controller Terminal Strip is correct
and secure.
2. If the buzzer difficulties continue,
contact RKI for further instruction.
Reset Switch
Not Working
•
The buzzer does not
silence when you
press the reset
switch.
•
The alarm 1 or
alarm 2 circuit does
not reset when you
press the reset
switch after the
alarm condition
passes.
•
The reset switch is
disconnected or
misconnected.
•
The reset switch is
malfunctioning.
1. Verify that the wiring to the
ALARM
RESET
(2) terminals of the Alarm/
Controller Terminal Strip is correct
and secure.
2. If the reset switch difficulties continue,
contact RKI for further instruction.
Strobe/Horn
Not Working
•
The strobe/horn
does not activate
when an alarm 1
condition occurs.
•
The strobe/horn is
malfunctioning.
•
The strobe/horn
wiring is disconnected
or misconnected.
1. Verify that the strobe/horn is wired
properly to the alarm terminal strip.
The red wire goes to the + terminal
and the black wire goes to the -
terminal.
2. If the strobe/horn difficulties continue,
contact RKI for further instruction.
Fail Condition
Rig Monitor indicates a
fail condition.
• The H
2
S detector
wiring is disconnected
or misconnected.
• The plug-in H
2
S
sensor is not properly
plugged into the
sockets in the detector
housing body.
• The detector’s zero
reading is low enough
to cause a fail
condition.
• The detector is
malfunctioning.
1. Verify that the H
2
S detector wiring is
correct and secure.
2. Confirm that the plug-in H
2
S sensor is
installed properly.
3. Perform a zero adjustment. A full
calibration is recommended.
4. If the fail condition continues, replace
the plug-in sensor as described later
in this section.
5. If the fail condition continues, contact
RKI for further instruction.
Table 7: Troubleshooting the Beacon 110 Rig Monitor (Continued)
Condition
Symptom(s)
Probable Causes
Recommended Action