Chapter 6: OTR 800 Description

6.1  OTDR Physical Description

Instrument Enclosure

The ODM® Brand OTDR 800 is packaged in a rugged housing and further protected by a rubberized boot. Although the front panel 

is weather-resistant, care must be taken to prevent liquids and contaminants from damaging the fragile optical and electrical 

connectors, as well as the glass display. Use a mild cleaning agent and soft damp cloth to clean up the panels and the outside 

case. Clean the optical connector if needed using only industry accepted fiber cleaning methods. NEVER open the instrument for 

cleaning. Return the unit to the factory for servicing, if necessary.

Front Panel

Top Panel

Fig 6.1

Power Button

Stylus

Color 

Touchscreen

Protective 

Rubber Boot

Fig 6.2

OTDR Port

Velcro 

Hand 

Strap

VFL Port

USB Charging / Data Port

Unit Serial Number

Battery Charge Indicator

6.2  Home Screen Display

The unit is equipped with a 3.5″ color TFT-resistive touch display.

ODM OTDR

CUSTOM 

TEST

VFL : OFF

   

 

VIEW TRACE

AUTO

TEST

Fig 6.3

AUTO TEST MODE

Tap this feature to run a simple 

Optical Time Domain Reflectometer test 

that quickly finds events and faults. 

VERSION & BATTERY INFORMATION

Shows version number for 5 seconds after start 

up and then battery gauge. 

CUSTOM TEST MODE

Tap this feature to run a more complex OTDR 

test. Designed for OTDR experts who need to set 

custom lengths, pulse widths, and IOR.

VIEW TRACE

Tap this icon to open the previously run 

trace results.

OTDR SETTINGS

Tap this icon to change the OTDR time, date, and 

sleep timer settings.

SAVED TESTS

Tap this icon to load previously saved tests.

VFL (VISUAL FAULT LOCATOR)

Tap this icon to toggle between the VFL 

modes, ON (CW), or 2HZ Modulation.

6.3  Power Requirements

The OTDR 800 is equipped with a 3500mAh Li-ion battery. The OTDR is supplied with a 100-240V USB power adapter with 5 VDC, 

2.1A output. A fully charged Li-ion battery will typically enable approximately 6 hours of use and require approximately 4 hours of 

recharging.

Kickstand

Содержание ODM OTDR 800

Страница 1: ...LITY IS ASSUMED BY RIPLEY TOOLS FOR ITS USE NOR FOR ANY INFRINGEMENTS OF PATENTS OR OTHER RIGHTS OF THIRD PARTIES THAT MAY RESULT FROM ITS USE NO LICENSE IS GRANTED BY IMPLICATION OR OTHERWISE UNDER A...

Страница 2: ...arning alerts to situations that could cause personal injury CAUTION A caution alerts to situations that may cause damage to the equipment or produce poor testing conditions resulting in inaccurate te...

Страница 3: ...e the AC adapter when there are signs of damage to the enclosure or cord Ensure that you are using the correct charger for the local line voltage Do not use any other charger than the one provided wit...

Страница 4: ...e range and pulse width Auto Test will set the averaging time to the best option dynamically typically 60 sec 3 The schematic view will be shown once the test has completed Review your results and or...

Страница 5: ...Display The unit is equipped with a 3 5 color TFT resistive touch display ODM OTDR CUSTOM TEST VFL OFF VIEW TRACE AUTO TEST Fig 6 3 AUTO TEST MODE Tap this feature to run a simple Optical Time Domain...

Страница 6: ...ok for the presence of a red spot on the paper 7 2 VFL Description The Visual Fault Locator emits visible red light at the 650 nm wavelength Its intended function is to allow an operator to identify t...

Страница 7: ...the wavelength list and choose the desired wavelength for the test Select either a single wavelength or two for dual mode operation Wavelengths are selected when they are highlighted in dark gray If t...

Страница 8: ...gth Once the first wavelength has finished the second test using the other wavelength will begin again at 0 Set Launch Receive Length Tap the textbox containing the fiber length to change it Enter the...

Страница 9: ...r the range of the values and set them using the same steps as for the test parameters NOTE There is an ORL and a Reflection Threshold setting The ORL is the return loss for the entire link system and...

Страница 10: ...ews low medium and high To set the event sensitivity tap the textbox for the sensitivity setting select the desired setting and then tap OK The lowest setting possible should be used to help filter ou...

Страница 11: ...l Parameters TRACE SETTINGS Shows Current Trace Parameters and Allows User to Edit Parameters for a New Test RUN RUN Runs a New Trace or Starts Live Mode Trace View Only TRACE VIEW Opens Trace View Sc...

Страница 12: ...ed end of the fiber under test The last row is with the type Link includes data that is relevant to the entire system link full fiber under test P or F Pass or Fail If any threshold is not met for an...

Страница 13: ...ss for the link is displayed System ORL 8 8 Event Analysis NOTE When using the Event Table or Schematic View it is necessary to keep in mind that Event Analysis provides approximate Loss and Distance...

Страница 14: ...available options Proceed to enter the rest of the test parameters as desired and then begin the test The two wavelength tests will run consecutively after one another After the first completes the s...

Страница 15: ...the display Set point number one by tapping the cursor select shown as an A or B letter in upper or lowercase refer to figure 8 8 button to toggle between the cursors until lowercase a is selected Mov...

Страница 16: ...roperly The green LSA Indicator Line overlays the backscatter of the Trace properly Chapter 9 Project Management 9 1 Project Management Description Project Management allows the user to save a set of...

Страница 17: ...reen Refer to Figure 9 1 Ensure that the right Project is highlighted because once it is deleted the project and all files associated with the project cannot be restored NOTE Delete must be held for a...

Страница 18: ...to the computer with the included USB cable NOTE The two plugs in the top right corner of the computer software screen should show as connected and green Highlight the file to be uploaded using the n...

Страница 19: ...nm 5 nm CONNECTOR TYPE 2 5 mm Universal OUTPUT POWER 1 mW Max LASER SAFETY CLASS II FDA21 CFR1040 10 1040 11 IEC 825 1 1993 Ripley Tools reserves the right to change specifications without notice Cha...

Страница 20: ...ode and Single Mode Fiber Construction of Fiber Optic Cables Connectors and Ferrule Types 2 Connector Inspection IEC 61300 3 35 Standard Understanding the IEC Specification Inspection Methods Connecto...

Отзывы: