81
Measurement
Settable measurement value of additional processing
The measurement value that can be set by the additional processing is different
according to the sub channel settings.
The relation between settable measurement value of additional processing
and the sub channel setting is as follows.
Measurement values of
additional processing
Sub channel settings
Measurement value
for each setting
Frequency weighting Time weighting
L
eq
A
I (Impulse)
L
AIeq
C
F (Fast)
L
Ceq
C
S (Slow)
L
max
A
I (Impulse)
L
AImax
L
peak
C
F (Fast)
L
Cpeak
C
S (Slow)
Z
F (Fast)
L
Zpeak
Z
S (Slow)
L
tm5
A
F (Fast)
L
Atm5
Note
If the additional processing value is not set to “ON”
on the display menu, measurement value are not
displayed.
When the optional Extended Function Program NX-
42EX is not installed, the time weighting characteristic
I (Impulse) cannot be selected.
Measurement value of additional processing is displayed
as
L
y within store data.