
37
Calibration
Before starting a measurement, the unit must be calibrated. There are two
types of calibration, namely electrical calibration using an internally generated
signal and acoustic calibration using an external sound calibrator.
Internal calibration (Electrical calibration)
Calibration is carried out using a signal generator (1 kHz, sinusoidal wave)
built into the unit.
1. Press the CAL key. A calibration screen such as shown below appears.
Verify the “Internal Calibration” is displayed in the upper part of
the screen.
If “Acoustic Calibration” is shown in the upper part of the screen,
press the DISPLAY key. The calibration mode will change to “In-
ternal Calibration”.
2. Confi rm that the calibration value indication shows 124 dB steadily.
If the bar graph upper limit setting is not 130 dB, a value of [Output
Level Range Upper −6 dB] will be fl ashing as the “124 dB” value
on the calibration value indication.
3. Use
the
keys to bring the level indication to the specifi ed value
(124.0 dB).
Содержание NL-42A
Страница 2: ......
Страница 4: ...ii...
Страница 8: ...vi...
Страница 10: ...viii...
Страница 18: ...xvi...
Страница 46: ...28 Preparations System screen Backlight LCD settings screen...
Страница 152: ...Unit mm Dimensional Drawings Bottom view Front view Side view Rear view 33 76 250 134 Specifications...
Страница 153: ......
Страница 154: ...No 66301 22 05...