33
Preparations
Comparator output
This is an open collector output that can be used to control external equipment.
When the optional Extended Function Program NX-42EX is not installed,
the comparator output cannot be set.
1. Press the MENU key to bring up the menu list screen.
2. Use
the
keys to select [I/O] and press the MENU/ENTER
key. The I/O screen appears.
3. Use the
keys to select [Comparator] and press the MENU/
ENTER key. The comparator screen appears.
4. Use the
keys to select [Comparator] and press the MENU/
ENTER key. The ON/OFF setting screen appears.
5. Use
the
keys to select [ON] and press the MENU/ENTER key.
6. Use
the
keys to select [Comparator level] and press the MENU/
ENTER key. The comparator level screen appears.
7. Use
the
keys to select the fi rst digit and use the
keys to
set the value.
8. Use the
keys to select the two lower digits and use the
keys to set the value. Then press the MENU/ENTER key. (Setting
range 25 dB to 130 dB, 1-dB steps)
9. Use
the
keys to select [Comparator band] and press the MENU/
ENTER key. The comparator band screen appears.
10. Use the
keys to select the comparator band (MAIN AP, SUB
AP) and press the MENU/ENTER key.
11. Press the START/STOP key to return to the measurement screen.
Содержание NL-42A
Страница 2: ......
Страница 4: ...ii...
Страница 8: ...vi...
Страница 10: ...viii...
Страница 18: ...xvi...
Страница 46: ...28 Preparations System screen Backlight LCD settings screen...
Страница 152: ...Unit mm Dimensional Drawings Bottom view Front view Side view Rear view 33 76 250 134 Specifications...
Страница 153: ......
Страница 154: ...No 66301 22 05...