Specifi cations
Maximum particle number concentration
40,000 particles/mL (coincidence loss is 10% or less)
False count rate
Average 0.1 particles/mL or less (measured with light
source off)
Warm-up time
Max. 15 minutes (after power-on)
Max. 5 minutes (from receiving laser-on command
in measurement pause/laser off condition, or relight
the light source after turning the light source off by
pressing the LASER OFF switch)
Indicators
Two color light emitting diode
PARTICLE MONITOR
Briefl y fl ashes green when particles of minimum de-
tectable particle size or above are detected
LIQUID LEAK
Lit (green) when leak is not detected within chassis
Lit (red) when leak is detected within chassis
CELL
Lit (green) during normal operation
Lit (red) when fl ow cell is contaminated, condensation
occurs or particle number concentration in sample fl uid
exceeded maximum particle number concentration
Off when light source is off
LASER
Lit (green) during normal operation
Flashing (green) when laser current has increased above
a certain threshold (light source nearing end of service
life; maintenance within 1 month recommended)
Lit (red) when light source temperature is out of range
Flashing (red) when light source output is not normal
Off when light source is off
POWER
Lit (green) while power to unit is on
Flashing (red) when internal program is not normal
Controls
LASER OFF switch
Changes ON/OFF of a light source by pressing for 2
seconds when the measurement is stopped
No effect when the measurement is in progress
No effect when there is a serious problem in the sensor