26
Measurement
When CELL indicator is lit in red
Status:
Measurement may not have been carried out properly, due
to excessive particle number concentration in sample fl uid,
fl ow cell contamination, condensation, or other cause.
Operation:
Measurement is possible, but may not be correct. Result
data should be used with care.
Output:
NOISE_HIGH (serial interface output only)
When LASER indicator is lit in red
Status:
Light source temperature is out of range.
Operation:
If measurement is in progress, it is stopped immediately
and measurement data are discarded.
Output: LASER_TEMP.
When LASER indicator fl ashes in red
Status:
Light source output has decreased to the rated level or
below, or becomes unstable.
Operation:
Measurement is possible, but may not be correct. Result
data should be used with care.
Output: LASER_FAIL
When LASER indicator fl ashes in green
Status:
Light source is nearing the end of its service life.
Operation:
Measurement is possible.
Output:
LASER_LIFE (serial interface output only)
When LASER indicator and CELL indicator are off
Status:
Light source is off.
Operation:
If measurement is in progress, it is stopped immediately
and measurement data are discarded.
Output:
LASER_OFF (serial interface output only)
_ represents a space.