27
Measurement
Automatic recovery of abnormal internal processing
If there is a problem with the internal processing for some reason, this unit
will automatically detect and recover it.
Detection of abnormal internal processing is done in measurement and at
the start of measurement.
Operation of each device associated with the recovery process is as follows.
Devices
Operation
KS-19F
Light source: turning on
(When the light source is lit in the measurement,
turn off the light source and turn on again.)
KE-40B1
Measurement
Measurement: stopped
Count value: not displayed
Setting of measurement
Measurement mode: manual
On Time, Off Time: 24:00
Alarm Size: 0.1
Alarm level: 0
Display
Pop-up screen: “SENSOR COMM. ERROR”
When KZ-30W1/W2 connected: “SENSOR
COMM. ERROR” and “SAMPLER STOPPED.”
During automatic recovery
Sensor model name of the title area: “--------”
LASER icon: not displayed
CELL icon: not displayed
Printer
Print: “!SENSOR COMM. ERROR”
Print of data: none
Serial interface
Information message of the measurement error
information request message “Q/E”:
Outputs “E/SENSOR COMM. ERROR”
Information message of the measurement data
information request message “Q/D”:
Outputs “D/”
Data store to
CF card
“Note”: write “Protected: SENSOR COMM. ERROR”
Particle count: space
KZ-30W1/W2
Plunger: going back to the home position
Operation mode, setup values: no change