22
Preparations
Particle size setting range
Using the controller KE-40B1, it is possible to freely specify the size ranges
for particle detection (up to 10 ranges).
Please note the following limitations with regard to setting particle size
ranges.
• Particle size ranges can be set to 0.03 to 0.13 μm in 0.01 μm steps.
• The upper limit for the smallest particle size channel (CH1) is 0.06
μm.
• It is necessary to set a particle size that is twice the value of the
smallest size range (CH1).
Note
If the setting a particle size that is twice the value
of the smallest size range (CH1) is not done, the unit
will set it automatically and will re-order the chan-
nels, starting from the smallest on CH1. Therefore
it may not be possible to set exactly the intended
range. The automatically set value (twice the CH1
range) cannot be deleted.