Remote Display Installation Manual
– Software Version 4.x
Page 44 D300-602-202
6.
Diagnostic Errors
The instrument continually monitors the condition of the
internal circuits. Any faults or out-of-tolerance conditions
are shown on the display as an
E
type error message.
In the table below the following terms are used:
Check
: This item can be checked on site by service
personnel.
Return for Service
: The instrument must be returned to
the manufacturer for factory service.
Error
Description
Resolution
(E0001)
The power supply
voltage is too low.
Check supply
(E0002)
The power supply
voltage is too high.
Check scale /
cables
(E0004)
Communications
data error (usually
self-correcting).
Check
transmission
format, baud rate,
stop bits & parity
(- - - - - -)
Communications
timeout (more than
1.5 seconds since
the last valid
communication from
the master, usually
self-correcting).
Check scale /
cables
The
E
type error messages are additive. For example if
power supply voltage is too low and a communications
data error occurs, the resulting error message will be
E0005
(0001 + 0004).
Содержание D32 series
Страница 1: ...D32x XD30 Remote Display Manual D300 601 200...
Страница 45: ...Remote Display Installation Manual Software Version 4 x D300 602 202 Page 45 Notes...
Страница 46: ...Remote Display Installation Manual Software Version 4 x Page 46 D300 602 202 Notes...
Страница 47: ...Remote Display Installation Manual Software Version 4 x D300 602 202 Page 47 Notes...
Страница 48: ...Remote Display Installation Manual Software Version 4 x Page 48 D300 602 202...