The following battery types are the only battery types that may be used.
1. Duracell Ultra MN1400
2. Panasonic LR14 (G)
3. Toshiba LR14
5−2.
Gas calibration
The measuring range of model RX-415 is dual expanded scale of 0-100%LEL and 0-100vol%
for HC, and 0-25vol% for O2. Then, the gas calibration for all ranges is required.
(1)
Prepare the calibration gas (both for low density gas and high density gas) and sampling bag.
(2)
Remove the filter tube and gas sampling tubes from the instrument. Then, arrange to connect
sampling bag from the gas inlet of instrument.
(3)
Turn on the power and make zero adjustment.
(4)
Put the calibration gas
*1
sampling bag and make gas calibration in the following procedure.
0.0%LEL 20.9%
(
Gas density display
)
↓
HOLD SPAN KEY
Press
PUMP
and
ZERO
↓
for 3 sec simultaneously.
APPLY GAS HC
Supply cal. gas from the gas
inlet and press
ENTER
When stop gas calibration, press
ESC
ENTER
ESC
To O2 calibration.
%LEL ADJ>−/+
Wait until gas reading gets stable.
When there is almost no difference between gas
display and calibration gas value, it does not need
gas calibration adjustment. Then press ESC switch
ESC
and make it end.
To O2 calibration.
%LEL ADJ>−/+
Adjust the gas reading to the calibration value.
By pressing
−
,
+
switch, adjust the gas reading to the
value of canned gas and press
ENTER
switch.
But when press
ENTER
without pressing
−
,
+
switch,
①
will appear. This means that calibration proceeds without
setting the calibration gas value.
When do not make gas calibration, press ESC
switch.
Gas cal. error
①
*
1:
Calibration gas
i−C
4
H
10
70〜80%LEL
Air-balance or N2-bal.