18
HC SPAN SET OK
Gas calibration is completed.
↓
APPLY GAS O2
Supply calibration gas from the gas inlet.
When perform gas calibration, press
ENTER
.
ESC
When stop gas calibration, press
ESC
.
Completion of gas cal.
0.0%LEL 20.9%
(
Gas density display
)
①
Display and function in gas calibration error. (See 3)
FAIL SPAN HC/
Danger
・
Do not approach the fire at the time of gas calibration at all.
For gas calibration, use the high density gas.
*
Note
・
Before gas calibration, confirm that the indication value is “0.0%LEL” after fresh air
is sucked. Be sure to make zero adjustment when the value is not “0.0%LEL”.
End even though “0.0%LEL” is displayed, if it is flickered, zero point is deviated.
Make zero adjustment.