20
(
6
)
Put the calibration gas
*3
to sampling bag and make gas calibration in the following procedure.
0.0%LEL 20.9%
(
Gas density display
)
↓
HOLD SPAN KEY
Press
PUMP
and
ZERO
↓
for 3 sec. simultaneously.
APPLY GAS HC
When perform gas calibration, press
ENTER
.
ESC
When stop gas calibration, press
ESC
.
To O2 calibration.
APPLY GAS O2
Supply calibration gas from the gas inlet
and press
ENTER
When stop gas calibration, press
ESC
.
ENTER
ESC
Gas calibration ends
→
Return to gas density display.
% ADJ>−/+
Wait until gas reading gets stable.
When there is almost no difference between gas
display and calibration gas value, it does not need
gas calibration adjustment. Then, press ESC switch
ESC
and make it end.
Gas calibration ends
→
Return to gas density display.
% ADJ>−/+
Adjust the gas reading to the calibration value.
By pressing
−
,
+
switch, adjust the gas reading to the
value of canned gas and press
ENTER
switch.
But when press
ENTER
without pressing
−
,
+
switch,
①
will appear. This means that calibration proceeds without
setting the calibration gas value.
When do not make gas calibration, press ESC switch.
Gas cal. error
②
O2 SPAN SET OK
Gas calibration is completed.
↓
0.0%LEL 20.9%
(
Gas density display
)
②
Display and function in gas calibration error.
FAIL SPAN /O2
*
3:
Calibration gas
N2 : Over 99.5vol%