. Calibration
7-1. Preparation for air and N2/CO2 calibrations
- 48 -
7
.
Calibration
7-1. Preparation
for air and N2/CO2 calibrations
Prepare the following before performing maintenance such as a bump test and air calibration.
Preparation for air calibration
Perform air calibration at maintenance before starting work or if the zero point deviates even though fresh
air is drawn. Before performing air calibration, check that the surrounding air is fresh.
・
Air calibration (P. 21, P. 55)
Preparation for N2/CO2 calibration
<Items to Prepare>
・
Calibration gas: N2 or CO2 of 100 vol% (recommended)
・
Gas sampling bag (*1)
・
Stopwatch
*1 Optional part
<Connection>
Connect the gas detector as shown in the figure below.
Connect a gas sampling bag at an appropriate timing.
Perform calibration in a single step using the concentration of the prepared calibration gas preliminarily
set to the gas detector, or perform it manually by adjusting to the concentration of the prepared calibration
gas.
・
AUTO CAL (P. 58)
・
ONE CAL (P. 60)
Gas sampling
bag
GAS OUT
Tube length: 300 mm