. Calibration
7-10. Password setting
- 64 -
7-10. Password setting
A password can be used to protect the entry to the calibration mode.
LCD display
1
Press the ENTER
button.
The gas detector enters the
password setting mode.
2
Select whether or not
to set a password
with the
▲/▼
button.
The state of <on> or <oFF>
for the current password
protection setting is
displayed.
* This is disabled <oFF> by
default.
3
Press the ENTER
button.
When <oFF> is confirmed,
the gas detector returns to the
calibration mode menu.
4
When <on> is
selected, use the
▲/▼
button to select a
value and press the
ENTER button.
The password is a four-digit
number. Set one digit at a
time.
▼