. Calibration
7-4. Air calibration
- 55 -
7-4. Air calibration
Perform the air calibration.
LCD display
1
Press the ENTER
button.
The gas detector enters the
air calibration mode.
2
Hold down the AIR
button.
3
Release the AIR
button when the
display changes from
"AirCL - HOLD AIR" to
"AdJ - RELEASE".
(Buzzer sound: Three times
<blip, blip, blip>)
After zero adjustment, "END"
is displayed and then the gas
detector returns to the
calibration mode menu.
(Buzzer sound: Once <blip>)
* If air calibration fails, "FAIL" is
displayed. Press the RESET
button to reset the alarm.