background image

Quick Start Guide

D e t e c t i o n

POLI Multi Gas Detector

Содержание POLI

Страница 1: ...Quick Start Guide D e t e c t i o n POLI Multi Gas Detector...

Страница 2: ...liser uniquement la batterie WatchGas PN M004 3002 W00 12 01 09 0001 WARNING Never operate the monitor when the cover is removed Remove the monitor cover only in area known to be non hazardous Use onl...

Страница 3: ...r ayant une concentration en oxyg ne sup rieure 21 La substitution de composants compromettra l aptitude la s curit intrins que et annulera la garantie Il est recommand de tester avec un gaz de concen...

Страница 4: ...icon on the LCD shows the charge status During active charging the icon cycles from empty to full and a yel low alarm light remains on which switches to green when completely charged Charging Comm Ca...

Страница 5: ...alarm function Two text and symbol boxes at the LCD bottom are mapped to the Left OK and Right keys The text or symbol in the boxes will vary with the menu Configuration Mode Hold down the Left and R...

Страница 6: ...ighlighted then press Left to exit Scrolling Numerical Entries While modifying numerical values such span values or alarm limits the function of the Left OK key can be switched between number increase...

Страница 7: ...and alarms functioning without doing a precise calibration To enter bump test Config mode Calibration Menu Bump Test Select sensors to be bump tested All sensors are selected by default Press Right t...

Страница 8: ...key during count down The Oxygen sensor is set to 20 9 during Fresh Air calibration and can be zeroed using the Nitrogen Calib menu Span Calibration Span calibration uses known concentrations of gas...

Страница 9: ...ion result Switch off the gas and exit Calib Mode Maintenance and Service For pumped units replace the inlet filter a regular basis especially when used in high dust environments For battery or sensor...

Отзывы: