background image

Turning the unit On

Press and hold the Down Key for 3 seconds, until the 

buzzer beeps and the red LED flashes. After a warm up 

and self-test sequence, the unit enters normal reading 

mode, showing real time gas concentrations.

Turning the unit Off

In normal reading mode, press and hold the Down 

Key for a 5-second count down, until “Unit Off” is 

displayed.

Charging

Always fully charge the battery before use. Plug the 

charger into the bottom of the POLI. The battery icon 

on the LCD shows the charge status. During active 

charging, the icon cycles from empty to full and a yel-

low alarm light remains on, which switches to green 

when completely charged.

Charging/Comm Cable

The USB Charging/Communications cable allows 

charging the unit on a personal computer. A local-

ly-purchased cable can work as well for charging, but 

WatchGas cable must be used for proper data transfer 

when communicating with WatchGas Suite.

Normal Mode

After warm-up is complete, the unit enters normal 

reading mode and displays the instantaneous gas 

concentrations continuously.inlet before the instru-

ment is used and replaced once its surface is dirty.

Содержание POLI

Страница 1: ...Quick Start Guide D e t e c t i o n POLI Multi Gas Detector...

Страница 2: ...liser uniquement la batterie WatchGas PN M004 3002 W00 12 01 09 0001 WARNING Never operate the monitor when the cover is removed Remove the monitor cover only in area known to be non hazardous Use onl...

Страница 3: ...r ayant une concentration en oxyg ne sup rieure 21 La substitution de composants compromettra l aptitude la s curit intrins que et annulera la garantie Il est recommand de tester avec un gaz de concen...

Страница 4: ...icon on the LCD shows the charge status During active charging the icon cycles from empty to full and a yel low alarm light remains on which switches to green when completely charged Charging Comm Ca...

Страница 5: ...alarm function Two text and symbol boxes at the LCD bottom are mapped to the Left OK and Right keys The text or symbol in the boxes will vary with the menu Configuration Mode Hold down the Left and R...

Страница 6: ...ighlighted then press Left to exit Scrolling Numerical Entries While modifying numerical values such span values or alarm limits the function of the Left OK key can be switched between number increase...

Страница 7: ...and alarms functioning without doing a precise calibration To enter bump test Config mode Calibration Menu Bump Test Select sensors to be bump tested All sensors are selected by default Press Right t...

Страница 8: ...key during count down The Oxygen sensor is set to 20 9 during Fresh Air calibration and can be zeroed using the Nitrogen Calib menu Span Calibration Span calibration uses known concentrations of gas...

Страница 9: ...ion result Switch off the gas and exit Calib Mode Maintenance and Service For pumped units replace the inlet filter a regular basis especially when used in high dust environments For battery or sensor...

Отзывы: