Product Specifications
<Disposal of Batteries>
Detection
principle
Galvanic cell type
Gas to be
detected *1
O2
Concentration
display
LCD digital display (seven-s
symbol)
Detection
range
0 to 40.0%
Display
resolution
0.1%
Detection
method
Diffusion type
Alarm
setpoint value
19.5%(1st)/23.5%LEL(2nd)/40%(OVER)
Displays
Battery level icon
Response
time (under
the same
conditions)
90% response: within 20 seconds
Gas alarm
type
2 gas alarm (1st/2nd), OVER
Gas alarm
display
Lamp blinking, intermittent buzzer
sounding, gas concentration display
blinking, and vibration
Gas alarm
activation
Self-latching (Non latching after reset)
Fault
alarm/self
diagnosis
System abnormalities, sensor
connection abnormalities, battery
voltage drop, calibration failure, and
clock abnormalities
Fault alarm
display
Lamp blinking, intermittent buzzer
sounding, and detail display
Functions
LCD backlight, peak display, and data
logger
Power supply AAA alkaline dry batteries x 2 (or AAA
Ni-MH batteries <Eneloop> available)
Continuous
operating
time
About 3000 hours (25
°
C, no alarm and
no lighting, alkaline dry batteries)
About 2000 hours (25
°
C, no alarm and
no lighting, Ni-MH batteries, batteries
fully charged)
Operating
temperature
-20 to +50ºC