24
6-2-3. Auto calibration
This is how to preset the concentration of the
prepared span calibration gas to the gas
monitor and perform calibration in a single
step.
Item LCD Details
A-CAL
POWER
Adjustment
Value Display
* Default setting 0.0%
* How to change
default setting.
First, press the AIR
+MODE Button.
Then, press AIR for
to change.
Finally, press MODE
for to enter.
POWER
A-CAL
(Blinking)
* Waiting for gas
introduction.
* Starts the gas
supply. After 60
seconds, press the
POWER button.
POWER
PASS
* After automatic
adjustment to the
preset value, the
monitor displays
PASS and then
returns to the user
mode menu.
* If span adjustment is
faulty, GAS NAME -
FAIL is displayed.
Press the MODE
button to reset.
A-CAL
* Stops the gas
supply.