25
6-2-4. Manual calibration
This is how to perform calibration with
manually set to the concentration value of the
prepared span calibration gas.
Item LCD Details
M-CAL
POWER
Concentration
Display
(Blinking)
* Waiting for gas
introduction.
* Press AIR + MODE
to decrease (or to
increase) the value.
AIR + MODE
* Starts the gas
supply. After 60
seconds, press the
AIR button to adjust
the value and then
press the POWER
button to confirm it.
POWER
END
* After displaying END,
the monitor returns to
the user mode menu.
* If span adjustment is
faulty, GAS NAME -
FAIL is displayed.
Press the MODE
button to reset.
M-CAL
* Stops the gas supply.