<Equipment required for gas calibration>
Span calibration gas
N2
99.9
vol%
or
more
(recommended)
Gas sampling bag
Polyurethane tube
Pump with the flow adjustment function
<adjustable to 500±10mL/min> (by pump,
flow meter or valve)
<Connection between
components>
Connect as shown in the
figure on the right and
adjust the flow.
Attach the gas sampling
bag whenever gas
introduction is required.
Stopwatch
Calibration cap (option) *1
*1 Use a dedicated calibration cap (option).
* The customer is required to provide all except for
*1.
Gas sampling
bag
Within a total tube length of 1 m Calibration cap
GAS IN
GAS OUT
Pump with the flow adjustment function
(by pump, flow meter or valve)
<Caution> Adjust the flow to 500±10mL/min.
<Caution> Leave GAS OUT open.