RIGOL
DS2000E User’s Guide
16-7
General Specifications
Probe Compensation Output
Output Voltage
[1]
About 3 V, peak-peak
Frequency
[1]
1 kHz
Power
Power Voltage
100 V to 240 V, 45 Hz to 440 Hz
Power
Maximum 50 W
Fuse
2 A, T Degree, 250 V
Environment
Temperature
Range
Operating: 0
℃
to +50
℃
Non-operating: -40
℃
to +70
℃
Cooling Method
Fan cooled
Humidity Range
0
℃
to +30
℃
:
≤95
%
RH
+30
℃
to +40
℃
:
≤75
%
RH
+40
℃
to +50
℃
:
≤45
%
RH
Altitude
Operating: under 3,000 m
Non-operating: under 15,000 m
Physical Characteristics
Dimensions
[4]
Width×Height×Depth=361.6 mm×179.6 mm×130.8 mm
Weight
[5]
Package Excluded
3.9 kg ± 0.2 kg
Package Included
4.5 kg ± 0.5 kg
Calibration Interval
The recommended calibration interval is 18 months.
Electromagnetic Compatibility and Safety
EMC
complies with EMC Directive 2014/30/EU,
complies with or above the standard specified in
IEC61326-1:2013/EN61326-1:2013 Group 1 Class A
CISPR 11/EN 55011
IEC 61000-4-2:2008/EN
61000-4-2
±
4.0 kV (contact discharge),
±
8.0
kV (air discharge)
IEC 61000-4-3:2002/EN
61000-4-3
3 V/m (80 MHz to 1 GHz); 3 V/m
(1.4 GHz to 2 GHz);
1 V/m (2.0 GHz to 2.7 GHz)
Содержание DS2000E Series
Страница 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 22: ......
Страница 62: ......
Страница 72: ......
Страница 82: ......
Страница 198: ......
Страница 212: ......
Страница 238: ......
Страница 258: ......
Страница 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Страница 268: ......
Страница 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...