RIGOL
Chapter 5 To Trigger the Oscilloscope
5-32
DS2000E User’s Guide
Duration Trigger (Option)
In duration trigger, the instrument identifies a trigger condition by looking for the
duration of a specified pattern. This pattern is a logical "AND" combination of the two
channels. Each channel can have a value of high (H), low (L) or don’t care (X). The
instrument triggers when the duration (
△
T) of this pattern meets the preset time, as
shown in the figure below.
T
T
CH1=H
CH2=L
CH1
CH2
Figure 5-11 Duration Trigger
Trigger Type:
Press Type to open the trigger type list. Rotate
and press down
to select
"Duration". At this point, the trigger setting information is displayed at the upper
right corner of the screen. For example,
. The trigger type is
duration trigger; the trigger source is CH1; the trigger level is 0.00 V.
Source Selection:
Press Source to open the signal source list and select CH1 or CH2. For the details,
please refer to the introduction in "Trigger Source". The current trigger source is
displayed at the upper right corner of the screen.
Pattern Setting:
Press Code to set the pattern of the current channel. At this point, the pattern
setting area (as shown in the figure below) is displayed at the bottom of the screen.
: set the pattern of the channel selected to "H", namely the voltage level is
higher than the trigger level of the channel.
: set the pattern of the channel selected to "L", namely the voltage level is
lower than the trigger level of the channel.
Содержание DS2000E Series
Страница 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 22: ......
Страница 62: ......
Страница 72: ......
Страница 82: ......
Страница 198: ......
Страница 212: ......
Страница 238: ......
Страница 258: ......
Страница 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Страница 268: ......
Страница 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...