Chapter 5 To Trigger the Oscilloscope
RIGOL
DS2000E User’s Guide
5-43
USB Trigger (Option)
Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the data packet on the
differential USB data cable (D+ and D-). This trigger supports USB Low Speed and
Full Speed. Below is the USB data transmission protocol diagram.
Figure 5-16 USB data transmission protocol diagram
Trigger Type:
Press Type to open the trigger type list. Rotate
and press down
to select
"USB". At this point, the trigger setting information is displayed at the upper right
corner of the screen. For example,
. The trigger type is USB trigger;
the trigger source is CH1; the trigger level is 0.00 V.
Source Selection:
Press D+ and D- to specify data sources for D+ and D- data cables respectively.
They can be set to CH1 or CH2. For the details, please refer to the introduction in
"
Trigger Source". The current trigger source is displayed at the upper right corner
of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Signal Speed:
Press Speed to select "Low Speed" (1.5 Mb/s) or "Full Speed" (12 Mb/s).
Trigger Condition:
Press When to select the desired trigger condition.
SOP: trigger at the sync bit
at the start of the data packet (SOP).
EOP: trigger at the end of the SEO portion of the EOP of the data packet.
RC: trigger when SEO is greater than 10 ms.
Suspend: trigger when the idle state of the bus is greater than 3 ms.
Содержание DS2000E Series
Страница 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 22: ......
Страница 62: ......
Страница 72: ......
Страница 82: ......
Страница 198: ......
Страница 212: ......
Страница 238: ......
Страница 258: ......
Страница 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Страница 268: ......
Страница 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...