RIGOL
Command Systems
Programming Guide for DS1000E, DS1000D Series
2-46
:TRIG:ALT:SLOP:LEVB ?
The query returns -1.500e+00.
15.
:TRIGger:ALTernation<mode>:COUPling
Command Format:
:TRIGger:ALTernation<mode>:COUPling
{DC|AC|HF|LF}
:TRIGger:ALTernation<mode>:COUPling?
Function Explanation:
The commands set and query the coupling mode. Thereinto,
DC: Allow all signals pass;
AC: Block DC signals and attenuate the signals lower than10Hz;
HF ( High frequency reject): Reject high frequency signals (Higher than 150
KHz);
LF ( Low frequency reject): Reject DC signals and attenuate low frequency
signals (Lower than 8KHz).
<mode>could be :EDGE, :PULSe or :SLOPe.
Returned Format:
The query returns DC , AC, HF or LF.
Example:
:TRIG:ALT:EDGE:COUP
DC Setup the coupling mode as DC.
:TRIG:ALT:EDGE:COUP?
The query returns DC.
16.
:TRIGger:ALTernation<mode>:HOLDoff
Command Format:
:TRIGger:ALTernation<mode>:HOLDoff <count>
:TRIGger:ALTernation<mode>:HOLDoff ?
Function Explanation:
The commands set and query the trigger holdoff. The <mode> could be
EDGE, :PULSe, :SLOPe or :VIDEO. The range of <count> is 500ns~1.5s.
Returned Format:
The query returns the setting value of <count> in s.
Example:
Содержание DS1000E Series, DS1000D Series
Страница 1: ...RIGOL Programming Guide DS1000E DS1000D Series Digital Oscilloscope July 2009 RIGOL Technologies Inc ...
Страница 2: ......
Страница 6: ......
Страница 86: ...RIGOL Command Systems Programming Guide for DS1000E DS1000D Series 2 74 LA GROU1 SIZ The query returns SMALL ...
Страница 102: ......
Страница 129: ...Programming Guide for DS1000E DS1000D Series 3 27 After you set the Start event see the result below ...
Страница 132: ......