Command Systems
RIGOL
Programming Guide for DS1000E, DS1000D Series
2-45
13.
:TRIGger:ALTernation:SLOPe:LEVelA
Command Format:
:TRIGger:ALTernation:SLOPe:LEVelA
<value>
:TRIGger:ALTernation:SLOPe:LEVelA?
Function Explanation:
The commands set and query the current upper boundary “Level A” of trigger
level in slope trigger mode. The range of <value> is LevelB ~ +6*Scale; Scale
indicates the current vertical scale, the unit is V/div.
Returned Format:
The query returns the setting value of <value> in V.
Example:
:TRIG:ALT:SLOP:LEVA
2 Setup the upper boundary “Level A” of
trigger level as 2V.
:TRIG:ALT:SLOP:LEVA?
The query returns 2.000e+00.
14.
:TRIGger:ALTernation:SLOPe:LEVelB
Command Format:
:TRIGger:ALTernation:SLOPe:LEVelB
<value>
:TRIGger:ALTernation:SLOPe:LEVelB?
Function Explanation:
The commands set and query the current lower boundary “Level B” of trigger
level in slope trigger mode. The range of <value> is -6*Scale ~ LevelA
,
Scale
indicates the current vertical scale, the unit is V/div.
Note:
The value of Level A (upper boundary) should not be less than the
maximum of Level B (lower boundary).
Returned Format:
The query returns the setting value of <value> in V.
Example:
:TRIG:ALT:SLOP:LEVB
-1.5
Setup the lower boundary “Level B” of
trigger level as -1.5V.
Содержание DS1000E Series, DS1000D Series
Страница 1: ...RIGOL Programming Guide DS1000E DS1000D Series Digital Oscilloscope July 2009 RIGOL Technologies Inc ...
Страница 2: ......
Страница 6: ......
Страница 86: ...RIGOL Command Systems Programming Guide for DS1000E DS1000D Series 2 74 LA GROU1 SIZ The query returns SMALL ...
Страница 102: ......
Страница 129: ...Programming Guide for DS1000E DS1000D Series 3 27 After you set the Start event see the result below ...
Страница 132: ......