Command Systems
RIGOL
Programming Guide for DS1000E, DS1000D Series
2-35
The query returns the setting value of level in V.
Note:
Level A (upper boundary) can not be less than the maximum of Level B
(lower boundary).
Example:
:TRIG:SLOP:LEVB
-1.5 Setup the lower boundary of trigger level as -1.5V.
:TRIG:SLOP:LEVB?
The query returns -1.500e+00.