8. Operating procedure
Connect the measuring heads to the controller. Apply supply voltage to the
controller. Values measured by channel 1 and channel 2 respectively are shown on digital
indicators. LED indicators show the position of the object in the measurement area, i.e. lit
areas and areas shaded by the object.
9. Working with parameters
Parameter control buttons and LED indicators are placed on the front panel of the
controller, Fig. 3.
Figure 3
Parametrization is performed for each channel separately. Press the
Channel
button to select the channel. Selected channel is indicated by the corresponding LED.
9.1. Settings menu
To enter the settings menu, you need to hold the
Menu
button pressed for 3
seconds.
9.2. Zero point
This parameter sets the measurement starting point within the measuring range.
After Zero Point is set, the measurement result (in the digital display and analog output) is
displayed as deviation from a zero point.
Press the
Channel
button and select the channel. The current channel number is
highlighted by LED. Install the controlled object to the required position within the range.
Enter the settings menu, select
"SEt.0"
using the arrows, and press the
Enter
button. Display 2 will show the current value. Select the current value of the parameter by
pressing the
Up
arrow, or select '3.50 mm' (the middle of the range) by pressing the
Down
arrow. Then press
Enter
to confirm selection.
To exit the settings menu, press the
Esc
button. You will be prompted to exit without
saving to FLASH by the
"CAnCEL"
message. To save parameters to FLASH, select
"SAUE"
and press
Enter
.