12
Edge Sensors. RF659 Series
RF659 [Revision 2.0.0] 26.06.2018
- Use the arrow buttons to set the parameter value so that the voltage at the analog
output AO1 is set to 5V.
- Press
Enter
to go to the next parameter.
- Display 1 shows the
"AOLo.2"
message. Display 2 shows the parameter value.
- Connect the voltmeter to terminals AO2 and AGnd.
- Use the arrow buttons to set the parameter value so that the voltage at the analog
output AO2 is set to -5V.
- Press
Enter
to go to the next parameter.
- Display 1 shows the
"AOHi.2"
message. Display 2 shows the parameter value.
- Use the arrow buttons to set the parameter value so that the voltage at the analog
output AO2 is set to 5V.
- Press
Enter
.
- To save the changes, select
"SAUE"
using the arrow buttons.
- To exit without saving, select
"CAnCL"
.
- Press
Enter
to exit the settings menu.
11. Controller settings for working with transparent
objects
When working with transparent objects, it may be necessary to perform additional
controller settings. This is due to the fact that the value of the threshold for finding the border
of transparent objects depends on the degree of their transparency.
The screenshots below show the signals from the CMOS sensor for various objects
and settings.
- Normal object.
- Transparent object (without correction, the
Level
parameter is not set).