RF659 [Revision 2.0.0] 26.06.2018
Edge Sensors. RF659 Series
11
9.7. Logic output thresholds
The controller is equipped with logical outputs (two logical outputs for each channel).
These logical outputs signalize about going beyond the upper and lower measurement
thresholds.
Press the
Channel
button and select the channel. The current channel number is
highlighted by LED.
Enter the menu, select
"LOUt.x"
using the arrows, and press the
Enter
button in
order to enter the submenu.
9.7.1. Lower threshold
The Threshold Low logical output is an isolated output of the open collector type (n-
p-n). It is set to the active state when the measurement value is lower than the value of the
lower limit parameter.
Indication of parameter number – the leftmost LED lights up on the LED line.
Display 1 shows the
"LO.x"
message, Display 2 shows the parameter value.
To edit the parameter, use the
Up
and
Down
buttons. To increase the parameter,
press
Up
. To decrease the parameter, press
Down
. In order to go to the next parameter,
press
Enter
.
9.7.2. Upper threshold
The Threshold High logical output is an isolated output of the open collector type (n-
p-n). It is set to the active state when the measurement value is higher than the value of the
upper limit parameter.
Indication of parameter number – the leftmost LED lights up on the LED line.
Display 1 shows the
"hI.x"
message, Display 2 shows the parameter value.
To edit the parameter, use the
Up
and
Down
buttons. To increase the parameter,
press
Up
. To decrease the parameter, press
Down
. In order to go to the main menu, press
Enter
.
9.8. Resetting parameters
Parameters can be reset to factory values. In this case, all parameters of both
channels will be reset.
Enter the menu, select
"dEFLt"
using the arrows, and press the
Enter
button in
order to enter the submenu.
To edit the parameter, use the
Up
and
Down
buttons. To set the default values,
press
Up
. Display 2 shows the
"LOAD"
message. To cancel the action, press
Down
.
Display 2 shows the
"CAnCEL"
message. In order to go to the main menu, press
Enter
.
10. Analog outputs
The analog outputs are set up once by the manufacturer, but in some cases an
adjustment may be required.
To enter the setting menu, it is necessary to do the following: when the controller is
turned off, press the
Down
button, turn on the power, and, holding the
Down
button
pressed, wait for the
"AOut.1"
and
"CLbr"
messages to appear.
- Connect the voltmeter to terminals AO1 and AGnd.
- Press the
Enter
button. Display 1 shows the
"AOLo.1"
message. Display 2
shows the parameter value.
- Use the arrow buttons to set the parameter value so that the voltage at the analog
output AO1 is set to -5V.
- Press
Enter
to go to the next parameter.
- Display 1 shows the
"AOHi.1"
message. Display 2 shows the parameter value.