12.2. Voltage output
The connection scheme is shown in the figure. To reduce noise, it is recom-
mended to install RC filter before the measuring instrument. The filter capacitor value is
indicated for maximum sampling frequency of the sensor (9,4 kHz) and this value in-
creases in proportion to the frequency reduction.
RF
60
5
Blue
Grey
10 kOm
5
n
F
Vo
ltmet
e
r
12.3. Configuration parameters
12.3.1. Range of the analog output.
While working with the analog output, resolution can be increased by using the
‗Window in the operating range‘ function which makes it possible to select a window of
required size and position in the operating range of the sensor within which the whole
range of analog output signal will be scaled.
If the beginning of the range of the analog signal is set at a higher value than the
end value of the range, this will change the direction of rise of the analog signal.
Note.
If the beginning of the range of the analog signal is set at a higher value
than the end value of the range, this will change the direction of rise of the analog signal.
12.3.2. Analog output operation mode.
When using ‗window in the operating range‘ function, this mode defines the ana-
log output operation mode.
Analog output can be:
in the window mode or
in the full mode.
"Window mode"
. The entire range of the analog output is scaled within the se-
lected window. Outside the window, the analog output is "0".
"Full mode"
. The entire range of the analog output is scaled within the selected
window (operating range). Outside the selected window, the whole range of the analog
output is automatically scaled onto the whole operating range of the sensor (sensitivity
range).
12.4. Factory parameters table
Range of the analog output
Measuring range of sensor
Analog output operation mode
Window