22

Attention!

Medical electrical equipment is subject to special precauti-

ons regarding electromagnetic compatibility (EMC).

Portable  and  mobile  radio  frequency  communication  de-

vices can affect medical electrical equipment. The ME de-

vice is for operation in an electromagnetic environment and 

intended for professional facilities such as industrial areas 

and hospitals.

The user of the device should ensure that it is operated wit-

hin such an en-vironment.

Warning!

The ME device may not be stacked, arranged or used directly 

next to or with other devices. When operation is required to 

be close to or stacked with other devices, the ME device and 

the other ME devices must be observed in order to ensure 

proper  operation  within  this  arrangement.  This  ME  device 

is intended for use by medical professionals only. This ME 

device is intended for use inProfessional Healthcare Facility 

Environment.  This  device  may  cause  radio  interference  or 

interfere  with  the  operation  of  nearby  devices.  It  may  be-

come  necessary  to  take  appropriate  corrective  measures, 

such as redirecting or rearranging the ME device or shield.

The rated ME device does not exhibit any basic performance 

features in the sense of IEC 60601-1, which would present 

an unacceptable risk to patients, operators or third parties 

should the power supply fail or malfunction.

Warning!

Portable RF communications equipment (radios) including 

accessories, such as antenna cables and external antennas, 

should not be used in closer proximity than 30 cm (12 inch-

es) to parts and cables of the Diagnostic station ri-former 

specified by the manufacturer. Failure to comply may result 

in a reduction in the device‘s performance features.

Directives and manufacturer‘s declaration - Electromagne-

tic emissions

The Diagnostic station ri-former are intended for use in the 

electromagnetic  environment  specified  below.  The  custo-

mer or user of the Diagnostic station ri-former should en-

sure that it is used in such an environment.

Guidance and manufacture’s declaration – electromagnetic emission

The 

ri-

former

®

 Diagnostic Station intended for use in the electromagnetic 

environment specified below. The customer of the user of the Diagnostic 

station ri-former should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic en-

vironment – guidance

RF-Emissionen

CISPR 11

Group 1

The 

ri-

former

®

  Dia-

gnostic  Station  uses 

RF  energy  only  for  its 

internal  function.  The-

refore,  its  RF  emissi-

ons  are  very  low  and 

are  not  likely  to  cause 

any interference in ne-

arby  electronic  equip-

ment.

RF-Emission

CISPR 11

Class B

The 

ri-

former

®

  Dia-

gnostic  Station  is  in-

tended  for  use  in  all 

establishments,  inclu-

ding  residential  areas 

and  those  directly 

connected  to  a  public 

supply  network  that 

also supplies buildings 

used  for  residential 

purposes. 

Harmonic emissions

IEC 61000-3-2

Pass

Voltage fluctuations/ 

flicker emissions

IEC 61000-3-3

Pass

Guidance and manufacture’s declaration – electromagnetic immunity

The 

ri-

former

®

 Diagnostic Station is intended for use in the electro-

magnetic environment specified below. The customer or the user of 

ri-

former

®

 Diagnostic Station should assure that it is used in such an 

environment.

Immunity test

IEC 60601 

test level

Compliance 

level

Electromagnetic 

environment - 

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

Electrical fast

transient/burst

IEC 61000-4-4

Con: ± 8 kV

Air: ± 15 kV

5/50 ns, 100 

kHz;       ±2 kV

Con: ± 8 kV

Air: ± 15 kV

5/50 ns, 100 kHz;       

±2 kV

Floors should be wood, 

concrete or ceramic 

tile. If floor are covered 

with synthetic material, 

the relative humidity 

should be at least 30%.

The quality of the 

supply voltage should 

be that of a typical 

business or hospital 

environment. 

Surge

IEC 61000-4-5

1.2/50 (8/20) µs

LtL:   ± 1.0 kV 

LtG:   ± 2.0 kV

1.2/50 (8/20) µs

LtL:   ± 1.0 kV 

LtG:   ± 2.0 kV

The quality of the 

supply voltage should 

be that of a typical 

business or hospital 

environment. 

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines

IEC 61000-4-11

0 % UT for 

0.5 cycle (1 

phase)

0 % UT for 1 

cycle

70 % UT for 

25/30 cycles 

(50/60 Hz)

0 % UT for 0.5 

cycle (1 phase)

0 % UT for 1 cycle

70 % UT for 25/30 

cycles (50/60 Hz)

The quality of the 

supply voltage should 

be that of a typical 

business or hospital 

environment.

0 % UT for 

250/300 

cycles (50/60 

Hz)

0 % UT for 

250/300 cycles 

(50/60 Hz)

Power frequency 

(50Hz/60Hz) magnetic 

field 

IEC 61000-4-8

30 A/m

50 Hz

30 A/m

50 Hz

Mains frequency ma-

gnetic fields should 

be at a level charac-

teristic of a typical 

location in a typical 

commercial hospital 

environment.

NOTE: U

T

 is the a.c. mains voltage prior to application of the test level.

Содержание ri-former

Страница 1: ...1 ri former Gebrauchsanweisung Instruction for use Instructions d utilisation Instrucciones de uso Istruzioni per l uso...

Страница 2: ...ie haben eine hochwertige Riester ri former Diagnosestation erwor ben die entsprechend der Richtlinie EU 2017 745 hergestellt wurde und st ndig strengsten Qualit tskontrollen unterliegt Bitte lesen Si...

Страница 3: ...tientenpopulation Das Ger t ist f r den Einsatz an allen Patienten vorgesehen 1 4 4 Bestimmungsgem e Anwender Benutzer Das Ger t kann von rzten und Pflegepersonal in Krankenh usern medizinischen Einri...

Страница 4: ...ebrauch die Integrit t und Vollst ndigkeit der ri former Diagnosestation des Erweiterungsmoduls der Instrumentenk pfe zu berpr fen Alle Komponenten m ssen miteinander kompatibel sein Inkompatible Komp...

Страница 5: ...Minuten ein 3 3 2 Entnehmen Sie den Handgriff nach oben aus der Griffhalterung und befestigen Sie den gew nschten Instrumentenkopf indem Sie die bei den hervorstehenden F hrungsnocken auf den Handgrif...

Страница 6: ...r t Medizinisches Ger t f r die Stromversor gung von Instrumenten Elektrischer Schutz Isolationsausr stung der Klasse II Modell ri former Diagnosestation mit ri former Erweiterungsmodul Netzteil Eing...

Страница 7: ...n der Patient bereits inner halb der letzten 24 Stunden mit diesem oder einem anderen ophthalmologischen Instrument untersucht wurde Dies gilt insbesondere dann wenn das Auge einer Netzhautfotogra fie...

Страница 8: ...abstehenden F hrungsnocken auf der Oberseite des Batteriegriffs aufsitzen Dr cken Sie den Instrumen tenkopf leicht auf den Batteriegriff und drehen Sie den Handgriff im Uhrzeigersinn bis zum Anschlag...

Страница 9: ...st und setzen Sie die Trichteraufnahme wieder auf 7 1 Otoskope L2 L3 ri derma Lampentr ger Nasenspekulum und Spatelhalter Entfernen Sie den Instrumentenkopf vom Batteriegriff Die Lampe be findet sich...

Страница 10: ...h ist sollte das ME Ger t beobachtet werden um seinen bestimmungsgem en Betrieb in dieser Anordnung zu berpr fen Dieses ME Ger t ist ausschlie lich f r den Gebrauch durch medizinisches Fachpersonal vo...

Страница 11: ...ngs stufe Elektromagnetische Umgebung Leitlinien HF Leitung IEC 61000 4 6 HF Strahlung IEC 61000 4 3 Umgebungsfelder von drahtlosen Funkkommunika tionsger ten Erf llt 3 V m 80 MHz bis 2 7 GHz Erf llt...

Страница 12: ...Stk ri former Instrumentenk pfe ohne Diebstahlsicherung mit Diebstahlsicherung ri scope L F O Otoskop Artikel Nr 10563 Artikel Nr 10563 301 L1 3 5 V XL L1 3 5 V XL mit Diebstahlsicherung Artikel Nr 10...

Страница 13: ...fter Handhabung Alle defekten Teile des Produkts werden inner halb der Garantiezeit kostenlos ersetzt oder repariert Dies gilt nicht f r Ver schlei teile F r das R1 Blutdruckmessger t sto fest gew hre...

Страница 14: ...tion 4 2 Cleaning and Disinfection 5 Technical Data 6 ri scope L Instrument heads ri scope Instrument heads 6 1 ri scope L otoscope 6 2 ri scope L ophthalmoscopes 6 3 Slit and spot retinoscopes 6 4 De...

Страница 15: ...4 3 Intended patient population The device is intended for all patients 1 4 4 Intended operators users The device could be used by a doctor nurse in hospitals medical facili ties clinics doctors offi...

Страница 16: ...All components must be compatible with each other Incompatible components can result in degraded perfor mance Never knowingly use a defective device Immediately replace parts that are broken worn miss...

Страница 17: ...bled ri former anesthesia device on the universal clamp and insert it Please ensure that both pins are introduced in the universal clamp Afterwards tighten the ri former anesthesia device with the lat...

Страница 18: ...t 2 2 x 5 V dc 2 x 1 15 A ri former Extension Module Input 5 V DC 3 A 15 W Output 1 1 x 3 5 V dc 700 mA Output 2 1 x 5 V dc 1 x 1 15 A Classification Application part type B Operating conditions 0 C t...

Страница 19: ...otography The light of this instrument may be harmful The risk of eye damage increases with the duration of irradiation An irradiation period with this instrument at maximum inten sity of longer than...

Страница 20: ...se direction For two modes of operation Fast expansion Push set screw on instrument head down with your thumb This setting does not allow changes in the position of the specula legs Individual expansi...

Страница 21: ...mp cloth until visual cleanliness is achieved Wipe with disinfectant according to the instructions of the disinfectant manufacturer Only cleaning agents with proven efficacy should be used under consi...

Страница 22: ...tion ri former should assure that it is used in such an environment Emission test Compliance Electromagnetic en vironment guidance RF Emissionen CISPR 11 Group 1 The ri former Dia gnostic Station uses...

Страница 23: ...or relocating the ri former Diagnostic Station b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended separation distances between portable and mobile RF c...

Страница 24: ...teVue head single LED 3 5 V with anti theft device for ri former wall station Art No 10513 EliteVue head single XL 3 5 V Art No 10513 301 EliteVue head single XL 3 5 V with anti theft device for ri fo...

Страница 25: ...Donn es techniques 6 T tes d instrument ri scope L T tes d instrument ri scope 6 1 Otoscope ri scope L 6 2 Ophtalmoscopes ri scope L 6 3 R tinoscopes fente et spot 6 4 Dermatoscope 6 5 Illuminateur br...

Страница 26: ...reil est destin tous les patients 1 4 4 Op rateurs utilisateurs pr vus L appareil peut tre utilis par un m decin ou une infirmi re dans des h pitaux des tablissements m dicaux des cliniques ou des cab...

Страница 27: ...rit et l int gralit de la station de diagnostic ri former du module d exten sion de la t te d instrument Tous les composants doi vent tre compatibles entre eux L incompatibilit des composants peut ent...

Страница 28: ...is de blocage Placez l appareil ri former anesth sie pr assembl sur la pince universelle et ins rez le Veuillez vous assurer que les deux broches sont introduites dans la pince universelle Serrez ensu...

Страница 29: ...60 Hz 0 6 A Sortie 5 V CC 3 A 15 W Station de diagnostic ri former Entr e 5 V CC 3 A 15 W Sortie 1 1 x 3 5 V CC 700 mA Sortie 2 2 x 5 V CC 2 x 1 15 A Module d extension ri former Entr e 5 V CC 3 A 15...

Страница 30: ...ologique au cours des 24 derni res heures Cela est particuli rement vrai lorsque l il a t expos une photographie r tinienne La lumi re de cet instrument peut tre dangereuse Le ris que de l sions ocula...

Страница 31: ...m illuminator LED 3 5 V 3 5 V 280 mA ave life 10 000 h 6 6 Sp culums nasaux 6 6 1 Utilisation pr vue Le sp culum nasal d crit dans ce mode d emploi est con u pour l c lairage et donc l examen de l int...

Страница 32: ...s d entretien 8 1 Remarque g n rale Le nettoyage et la d sinfection des instruments m dicaux servent prot ger le patient l utilisateur et les tiers et conserver les instru ments m dicaux en bon tat En...

Страница 33: ...i dessous Le client ou l utilisateur de la station de diagno stic ri former doit s assurer que celle ci est utilis e dans un tel environnement Directives et d claration du fabricant missions lectromag...

Страница 34: ...es telles que la r orientation ou le d placement de la station de diagnostic ri former b Sur la plage de fr quence de 150 kHz 80 MHz les intensit s de champ doivent tre inf rieures 3 V m Distances de...

Страница 35: ...eVue simple XL 3 5 V Art n 10513 301 T te EliteVue simple XL 3 5 V avec dispositif antivol pour la station murale ri former 12 limination Attention Les dispositifs m dicaux utilis s doivent tre jet s...

Страница 36: ...fecci n 5 Datos t cnicos 6 Cabezales de instrumento ri scope L Cabezales de instrumento ri scope 6 1 Otoscopio ri scope L 6 2 Oftalmoscopio ri scope L 6 3 Retinoscopios de hendidura y punto 6 4 Dermat...

Страница 37: ...os los pacientes 1 4 4 Operadores usuarios previstos El dispositivo puede ser utilizado por personal m dico o de enfermer a en hospitales centros m dicos cl nicas o consultorios m dicos No se puede us...

Страница 38: ...usuario Es su responsabilidad Antes de cada uso el usuario debe verificar la integridad y la totalidad de la unidad de diagn stico ri former m du lo de extensi n cabezal de instrumentos Todos los com...

Страница 39: ...tivo de an estesia ri former premontado en la abrazadera universal e ins rtelo Aseg rese de que ambos pasadores se introducen en la abrazadera universal Luego apriete el dispositivo de anestesia ri fo...

Страница 40: ...1 x 3 5 Vcc 700 mA Salida 2 2 x 5 Vcc 2 x 1 15 A M dulo de extensi n ri former Entrada 5 Vcc 3A 15 W Salida 1 1 x 3 5 Vcc 700 mA Salida 2 1 x 5 Vcc 1 x 1 15 A Clasificaci n Componente de aplicaci n ti...

Страница 41: ...as Esto es especialmente cierto cuando el ojo ha sido expuesto a la fotograf a retiniana La luz de este instrumento puede ser da ina El riesgo de da o ocular aumenta con la duraci n de la irradiaci n...

Страница 42: ...V 3 5 V 280 mA promedio de vida til 10 000 h 6 6 Esp culo nasal 6 6 1 Uso previsto El esp culo nasal descrito en estas instrucciones de funcionamiento ha sido fabricado para iluminar y por lo tanto ex...

Страница 43: ...o en el cabezal del instrumento para que la base de la l mpara encaje en la ranura del cabezal del instrumento 8 Instrucciones de conservaci n 8 1 Nota general La limpieza y desinfecci n de los dispos...

Страница 44: ...da para usarse en el entorno electromagn tico especificado a conti nuaci n El cliente o el usuario de la unidad de diagn stico ri former debe garantizar que se use en dicho entorno Gu a y declaraci n...

Страница 45: ...ubicaci n de la unidad de diagn stico ri former b En el rango de frecuencias de 150 kHz a 80 MHz las intensidades de campo deben ser inferiores a 3 V m Distancias de separaci n recomendadas entre equi...

Страница 46: ...LED 3 5 V LED dispositivo antirrobo EliteVue N de art culo 10512 Cabezal EliteVue simple LED 3 5 V N de art culo 10512 301 Cabezal EliteVue simple LED 3 5 V con dispositivo antirrobo para estaci n de...

Страница 47: ...1 Informazioni generali 4 2 Pulizia e disinfezione 5 Dati tecnici 6 Teste strumentali ri scope L Teste strumentali ri scope 6 1 Otoscopio ri scope L 6 2 Oftalmoscopi ri scope L 6 3 Retinoscopi a fessu...

Страница 48: ...spositivo pu essere utilizzato da medici e personale infermieristi co presso ospedali strutture sanitarie cliniche e ambulatori vietato l uso negli ambienti RM 1 4 5 Competenze formazione richieste ag...

Страница 49: ...ture sanitarie cliniche e ambulatori Attenzione Responsabilit dell utente responsabilit dell utente Prima di ogni utilizzo verificare l integrit e la comple tezza della Stazione diagnostica ri former...

Страница 50: ...universale e inserirlo Assicurarsi che entrambi i perni siano inseriti nel morsetto universale Fissare quindi il dispositivo ri former anestesia serrando la vite laterale 3 1 3 Fissaggio delle piastr...

Страница 51: ...resso 5 VDC 3A 15 W Uscita 1 1 x 3 5 V dc 700 mA Uscita 2 2 x 5 V dc 2 x 1 15 A Modulo di estensione ri former Ingresso 5 VDC 3A 15 W Uscita 1 1 x 3 5 Vdc 700 mA Uscita 2 1 x 5 Vdc 1 x 1 15 A Classifi...

Страница 52: ...oporzionale alla durata dell irraggiamento Il valore standard del rischio viene superato quando la durata dell irradiazione luminosa proveniente da questo strumento superiore ai 5 minuti Nonostante no...

Страница 53: ...a bat terie e ruotarlo in senso orario fino alla posizione finale Rimuovere la testa ruotando in senso antiorario Per due modalit operative Divaricazione rapida Spingere con il pollice la vite di rego...

Страница 54: ...ali e dei manici Pulire con disinfettante seguendo le istruzioni del produttore del disin fettante Devono essere utilizzati solo disinfettanti di comprovata effica cia in considerazione delle linee gu...

Страница 55: ...io ne diagnostica ri former devono assicurarsi che sia utilizzata in ambienti corrispondenti Test delle emissioni Conformit Ambiente elettromagnetico guida Emisiones de radiofrecuencia CISPR 11 Gruppo...

Страница 56: ...ssarie misure aggiuntive quali il riorientamento o il riposizionamento della Stazione diagnostica ri former b Oltre la gamma di frequenza da 150 kHz a 80 MHz le intensit di campo devono essere inferio...

Страница 57: ...3 5 V Art n n 10512 301 Testa EliteVue singola LED 3 5 V con dispositivo anti furto per stazione diagnostica a parete ri former Art n n 10513 Testa EliteVue singola XL 3 5 V Art n n 10513 301 Testa E...

Страница 58: ...3 1 4 4 1 4 5 1 4 6 1 5 1 6 2 2 1 2 2 3 3 1 3 2 3 3 3 4 ri former Mobile 3 5 ri former Anesthesia 4 4 1 4 2 5 6 ri scope L ri scope 6 1 ri scope L 6 2 ri scope L 6 3 6 4 6 5 6 6 6 7 6 8 6 9 6 10 7 L1...

Страница 59: ...59 1 3 IEC 60601 1 2 1 4 ri former 1 4 1 ri former C F 2002 96 EC Max 1 5 kg Mass 13 kg ri former big ben 1 4 2 1 4 3 1 4 4 1 4 5 1 4 6 1 5 ri former 2...

Страница 60: ...ri former ri former ri former ri former Ries ter 3650 ri former 1 3 5 100 240 3652 3652 ri former 2 3 5 100 240 3650 300 ri former 2 3 5 100 240 3652 300 ri former 2 3 5 100 240 ri former ri former 1...

Страница 61: ...2 3 3 3 3 1 HR MIN 3 3 2 3 3 3 100 3 3 4 rheotronic 3 4 ri former Mobile ri former 3 5 ri former Anesthesia ri former An esthesia ri former Anesthesia 3 1 3 3 1 4 a b c 3 1 5 b c I ON 0 OFF HR MIN B 3...

Страница 62: ...ormer 100 240 50 60 0 6 5 3 15 ri former 5 3 15 1 1 3 5 700 2 2 5 2 1 15 ri former 5 3 15 1 1 3 5 700 2 1 5 1 1 15 B 0 C 40 C 10 85 6 ri scope L ri scope ri scope L ri scope L b a b a 5 C 50 C 10 85 7...

Страница 63: ...3 6 1 3 360 6 1 4 180 3 6 1 5 6 1 6 XL 2 5 2 5 750 15 XL 3 5 3 5 720 15 LED 2 5 2 5 280 10 000 LED 3 5 3 5 280 10 000 6 2 ri scope L 6 2 1 Riester 24 5 DIN EN 62471 2 3 6 2 2 L1 L2 1 10 12 15 20 40 1...

Страница 64: ...ma 6 4 1 ri derma 6 4 2 6 4 3 6 4 4 2 0 10 6 4 5 ri derma XL 2 5 2 5 750 15 ri derma XL 3 5 3 5 690 15 ri derma LED 2 5 2 5 280 10 000 ri derma LED 3 5 3 5 280 10 000 24 6 5 6 5 1 6 5 2 6 5 3 XL 2 5 2...

Страница 65: ...5 280 10 000 6 8 6 8 1 Riester 6 8 2 6 9 6 9 1 Riester 6 9 2 6 9 3 360 2 5 6 9 4 6 9 5 HL 2 5 2 5 680 20 XL 3 5 3 5 700 20 6 10 6 10 1 Riester 6 10 2 6 10 3 2 5 360 6 10 4 6 10 5 HL 2 5 2 5 680 40 XL...

Страница 66: ...1 5 IEC 60601 1 30 12 ri former ri former ri former ri former ri former CISPR 11 1 ri former CISPR 11 B La Stazione diagnostica ri for mer idonea all uso in tutti gli ambienti compresi gli ambienti re...

Страница 67: ...2 0 IEC 61000 4 11 0 UT 0 5 1 0 UT 1 70 UT 25 30 50 60 0 UT 0 5 1 0 UT 1 70 UT 25 30 50 60 0 UT 250 300 50 60 0 UT 250 300 50 60 50 60 IEC 61000 4 8 30 A 50 30 A 50 UT ri former ri former IEC 60601 IE...

Страница 68: ...scope L 10537 10537 301 3 5 V XL 3 5 V XL 10575 10575 301 3 5 V LED 3 5 V LED ri scope L portalampadina a fibra ottica 10539 10539 301 3 5 V XL 3 5 V XL 10576 10576 301 3 5 V LED 3 5 V LED EliteVue 10...

Страница 69: ...69 12 2 B R1 shock proof 5 CE P Riester Rudolf Riester GmbH Dept Repairs RR Bruckstr 31 72417 Jungingen Germany...

Страница 70: ...70...

Страница 71: ...71...

Страница 72: ...417 Jungingen Deutschland Tel 49 7477 9270 0 Fax 49 7477 9270 70 info riester de www riester de 99249 Rev 0_2020 05 nderungen vorbehalten Subject to alterations Sous reserve de modifications Sujeto a...

Отзывы: